FDA Enforcement Class II Terminated

SIEMENS IMMULITE/IMMULITE 2000 Systems Rubella IgM (RUM), REF/Catalog Number for US: L2KRM2(D), Siemens Material Number (SMN) for US: 10374021, REF/Catalog # for OUS: L2KRM2, SMN for OUS: 10381327; Unique Device Identification Number (UDI) 00630414961996; IVD --- For in vitro diagnostic use with the IMMULITE/IMMULITE 1000/IMMULITE 2000/IMMULITE 2000 XPi Systems Analyzers-for the qualitative detection of IgM antibodies to Rubella virus in human serum or plasma (EDTA and heparinized), as an aid in the presumptive diagnosis of an acute or recent rubella infection, particularly in women of childbearing age

Recall: Z-0082-2017 · Reported October 19, 2016

Enforcement

Recall Number
Z-0082-2017
Event ID
75037
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Healthcare Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 19, 2016
Initiation Date
August 24, 2016
Classification Date
October 13, 2016
Termination Date
May 24, 2018
Address
511 Benedict Ave, N/A, Tarrytown, NY, 10591-5005, United States

Description

SIEMENS IMMULITE/IMMULITE 2000 Systems Rubella IgM (RUM), REF/Catalog Number for US: L2KRM2(D), Siemens Material Number (SMN) for US: 10374021, REF/Catalog # for OUS: L2KRM2, SMN for OUS: 10381327; Unique Device Identification Number (UDI) 00630414961996; IVD --- For in vitro diagnostic use with the IMMULITE/IMMULITE 1000/IMMULITE 2000/IMMULITE 2000 XPi Systems Analyzers-for the qualitative detection of IgM antibodies to Rubella virus in human serum or plasma (EDTA and heparinized), as an aid in the presumptive diagnosis of an acute or recent rubella infection, particularly in women of childbearing age

Reason

Siemens received multiple customer complaints indicating an increase in false positive and indeterminate patient sample results and increased imprecision for several lots of Rubella IgM assay on the IMMMULITE 2000/IMMULITE 2000 XPi platforms.

Code Info

Kit lots 236M, 237, 238, 239, 239L (OUS) and D236M and D238 (US)

Distribution

Worldwide Distribution n- US (Nationwide) and Foreign distribution to the following countries: Argentina, Bangladesh, Brazil, Bulgaria, China, Colombia, Costa Rica, Cyprus, Egypt, France, Germany, Greece, Guatemala, India, Indonesia, Israel, Italy, Kazakhstan, Macedonia, Mexico, Paraguay, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, South Africa, Syria, United Arab Emirates, and Uruguay.

Quantity

Domestic: 177 Kits; Foreign: 3435 Kits