FDA Enforcement Class II Ongoing

Medfusion Syringe Pump, Model 4000-XXXX-XX: a) 0100-50, b) 0100-249, c) 0101-50, d) 0101-51, e) 0101-78, f) 0101-249, g) 0105-51, h) 0105-78, i) 0105-249, j) 0106-00, k) 0106-01, l) 0106-231, m) 0107-01, software versions: v3, v4, v5, and v6

Recall: Z-0079-2024 · Reported October 18, 2023

Enforcement

Recall Number
Z-0079-2024
Event ID
92956
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Smiths Medical ASD Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 18, 2023
Initiation Date
August 16, 2023
Classification Date
October 6, 2023
Address
6000 Nathan Ln N, N/A, Minneapolis, MN, 55442-1690, United States

Description

Medfusion Syringe Pump, Model 4000-XXXX-XX: a) 0100-50, b) 0100-249, c) 0101-50, d) 0101-51, e) 0101-78, f) 0101-249, g) 0105-51, h) 0105-78, i) 0105-249, j) 0106-00, k) 0106-01, l) 0106-231, m) 0107-01, software versions: v3, v4, v5, and v6

Reason

A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will display a System Failure Alarm (including Force Sensor BGND Test, Force Sensor Bridge Test, or Force Sensor Test). However, if the calibration shift is not large enough to trigger a System Failure Alarm, the threshold to detect an occlusion may increase, increasing the time to occlusion detection, or the threshold may decrease, leading to false occlusion alarms. Although shifts in the force sensor calibration may occur over time with any device, an increased potential for such shifts has been reported in devices produced before April 2022 due to mechanical interference between parts of the plunger head assembly. Out of an abundance of caution, we are notifying all customers of this potential issue.

Code Info

a) 0100-50, UDI/DI 10610586035326; b) 0100-249, UDI/DI 10610586033124; c) 0101-50, UDI/DI 10610586040054; d) 0101-51, UDI/DI 10610586040818; e) 0101-78, UDI/DI 10610586040788; f) 0101-249, UDI/DI 10610586040078; g) 0105-51, UDI/DI 10610586043567; h) 0105-78, UDI/DI 10610586043574; i) 0105-249, UDI/DI 10610586043581; j) 0106-00, UDI/DI 15019517070750; k) 0106-01, UDI/DI 15019517069624; l) 0106-231, UDI/DI 15019517070996; m) 0107-01, UDI/DI 15019517154313; All serial numbers

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

Quantity

67507 pumps