FDA Enforcement
Class II
Terminated
Radiation Therapy Treatment Planning System, Model 5.0
Recall: Z-0079-2017
·
Reported October 19, 2016
Enforcement
- Recall Number
- Z-0079-2017
- Event ID
- 74750
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- RAYSEARCH LABORATORIES AB
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 19, 2016
- Initiation Date
- May 30, 2016
- Classification Date
- October 13, 2016
- Termination Date
- September 22, 2017
- Address
- Sveavaegen 9, N/A, Stockholm, N/A, N/A, Sweden
Description
Radiation Therapy Treatment Planning System, Model 5.0
Reason
For a treatment plan consisting of multiple beam sets, the table for ROI plan dose statistics in the report may show the statistics for a beam set dose. The error can only be triggered when using a report template where statistics for a beam set dose is included as the last dose statistics prior to the plan dose statistics.
Code Info
5.0.0.37, 5.0.1.11 and 5.0.2.35
Distribution
AZ, CA, FL, MT, NC NY, TX, OH & WA
Quantity
491 units (187 domestically & 304 internationally)