FDA Enforcement Class II Terminated

Radiation Therapy Treatment Planning System, Model 5.0

Recall: Z-0079-2017 · Reported October 19, 2016

Enforcement

Recall Number
Z-0079-2017
Event ID
74750
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
RAYSEARCH LABORATORIES AB
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 19, 2016
Initiation Date
May 30, 2016
Classification Date
October 13, 2016
Termination Date
September 22, 2017
Address
Sveavaegen 9, N/A, Stockholm, N/A, N/A, Sweden

Description

Radiation Therapy Treatment Planning System, Model 5.0

Reason

For a treatment plan consisting of multiple beam sets, the table for ROI plan dose statistics in the report may show the statistics for a beam set dose. The error can only be triggered when using a report template where statistics for a beam set dose is included as the last dose statistics prior to the plan dose statistics.

Code Info

5.0.0.37, 5.0.1.11 and 5.0.2.35

Distribution

AZ, CA, FL, MT, NC NY, TX, OH & WA

Quantity

491 units (187 domestically & 304 internationally)