FDA Enforcement Class II Terminated

Electrohydraulic lithotripter probe, 1.9 FR 375 cm, Model 9-195-371DS used with a Model 9-201-00 Autolith lithotripter. electrohydraulic lithotripsy (EHL) is a method that allows large stones to be fragmented and removed percutaneously or endoscopically.

Recall: Z-0073-2018 · Reported November 15, 2017

Enforcement

Recall Number
Z-0073-2018
Event ID
77367
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Northgate Technologies, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 15, 2017
Initiation Date
May 24, 2017
Classification Date
November 8, 2017
Termination Date
June 9, 2020
Address
1591 Scottsdale Ct, N/A, Elgin, IL, 60123-9361, United States

Description

Electrohydraulic lithotripter probe, 1.9 FR 375 cm, Model 9-195-371DS used with a Model 9-201-00 Autolith lithotripter. electrohydraulic lithotripsy (EHL) is a method that allows large stones to be fragmented and removed percutaneously or endoscopically.

Reason

The inner and outer labels on the product contain incorrect reference number of 9-202-3751DS. According to Northgate Technologies, the resistor in the probe will not allow the mislabeled product to be used with the incorrect generator. Could Delay Care.

Code Info

Model No. 9-195-371DS, Lot/Serial Number LDS12516

Distribution

US Distribution to the following states : MA, NE and TN and Internationally to Canada.

Quantity

186