Automated Impella Controller (AIC), used for left heart support blood pump, labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, AU; Product Code: 0042-0000-AU. 2. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 3. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 4. Impella Controller, Packaged, EU, Loaner; Product Code: 0042-0000-EU-L. 5. Impella Controller, Packaged, IN; Product Code: 0042-0000-IN. 6. Impella Controller, Packaged, JP; Product Code: 0042-0000-JP. 7. Impella Controller, Japan, Loaner; Product Code: 0042-0000-JP-L. 8. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 9. Impella Controller, UK, Loaner; Product Code: 0042-0000-UK-L. 10. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 11. Impella Controller, Packaged, US, Loaner; Product Code: 0042-0000-US-L. 12. Impella Optical Controller, Packaged, AU; Product Code: 0042-0010-AU. 13. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 14. Impella Optical Controller, EU, Loaner; Product Code: 0042-0010-EU-L. 15. Impella Optical Controller, Packaged, IN; Product Code: 0042-0010-IN. 16. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 17. Impella Optical Controller, UK, Loaner; Product Code: 0042-0010-UK-L. 18. Impella Optical Controller, Packaged, US; Product Code: 0042-0010-US. 19. Impella Optical Controller, US, Loaner; Product Code: 0042-0010-US-L. 20. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU. 21. Optical AIC w/Impella Connect,Pack'd, CA; Product Code: 0042-0040-CA. 22. AIC w/Remote Link, Pkgd, Cln Trl, US; Product Code: 0042-0040-CT. 23. AIC w/Remote Link, Pkgd, Cln Trl, US; Product Code: 0042-0040-CT-L. 24. Optical AIC w/Impella Connect,Pack'd, EU; Product Code: 0042-0040-EU. 25. Optical AIC w/Impella Connect,EU,Loaner; Product Code: 0042-0040-EU-L. 26. Optical AIC w/Impella Connect, Pkgd, JP; Product Code: 0042-0040-JP. 27. Optical AIC w/Impella Connect, Pkgd, JP; Product Code: 0042-0040-JP-L. 28. Optical, AIC, Impella Connect, Pkgd, UK; Product Code: 0042-0040-UK. 29. Optical, AIC, Impella Connect, Pkgd, US; Product Code: 0042-0040-US. 30. Optical AIC Impella Connect, US, Loaner; Product Code: 0042-0040-US-L. 31. Dbl optical, AIC Impella Connect, Phg US; Product Code: 1000201. 32. AIC w/Impella Connect for ECP; Product Code: 1000432.
Enforcement
- Recall Number
- Z-0069-2026
- Event ID
- 97617
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Abiomed, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 15, 2025
- Initiation Date
- October 1, 2025
- Classification Date
- October 9, 2025
- Address
- 22 Cherry Hill Dr, Danvers, MA, 01923-2575, United States
Description
Automated Impella Controller (AIC), used for left heart support blood pump, labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, AU; Product Code: 0042-0000-AU. 2. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 3. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 4. Impella Controller, Packaged, EU, Loaner; Product Code: 0042-0000-EU-L. 5. Impella Controller, Packaged, IN; Product Code: 0042-0000-IN. 6. Impella Controller, Packaged, JP; Product Code: 0042-0000-JP. 7. Impella Controller, Japan, Loaner; Product Code: 0042-0000-JP-L. 8. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 9. Impella Controller, UK, Loaner; Product Code: 0042-0000-UK-L. 10. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 11. Impella Controller, Packaged, US, Loaner; Product Code: 0042-0000-US-L. 12. Impella Optical Controller, Packaged, AU; Product Code: 0042-0010-AU. 13. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 14. Impella Optical Controller, EU, Loaner; Product Code: 0042-0010-EU-L. 15. Impella Optical Controller, Packaged, IN; Product Code: 0042-0010-IN. 16. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 17. Impella Optical Controller, UK, Loaner; Product Code: 0042-0010-UK-L. 18. Impella Optical Controller, Packaged, US; Product Code: 0042-0010-US. 19. Impella Optical Controller, US, Loaner; Product Code: 0042-0010-US-L. 20. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU. 21. Optical AIC w/Impella Connect,Pack'd, CA; Product Code: 0042-0040-CA. 22. AIC w/Remote Link, Pkgd, Cln Trl, US; Product Code: 0042-0040-CT. 23. AIC w/Remote Link, Pkgd, Cln Trl, US; Product Code: 0042-0040-CT-L. 24. Optical AIC w/Impella Connect,Pack'd, EU; Product Code: 0042-0040-EU. 25. Optical AIC w/Impella Connect,EU,Loaner; Product Code: 0042-0040-EU-L. 26. Optical AIC w/Impella Connect, Pkgd, JP; Product Code: 0042-0040-JP. 27. Optical AIC w/Impella Connect, Pkgd, JP; Product Code: 0042-0040-JP-L. 28. Optical, AIC, Impella Connect, Pkgd, UK; Product Code: 0042-0040-UK. 29. Optical, AIC, Impella Connect, Pkgd, US; Product Code: 0042-0040-US. 30. Optical AIC Impella Connect, US, Loaner; Product Code: 0042-0040-US-L. 31. Dbl optical, AIC Impella Connect, Phg US; Product Code: 1000201. 32. AIC w/Impella Connect for ECP; Product Code: 1000432.
Potential cybersecurity vulnerabilities related to the operating system in the Automated Impella Controller (AIC).
1. Product Code: 0042-0000-AU; GTIN: 813502010886; All Serial Numbers. 2. Product Code: 0042-0000-CA; GTIN: 813502011272; All Serial Numbers. 3. Product Code: 0042-0000-EU; GTIN: 813502011289; All Serial Numbers. 4. Product Code: 0042-0000-EU-L; GTIN: 813502011289; All Serial Numbers. 5. Product Code: 0042-0000-IN; GTIN: 813502011418; All Serial Numbers. 6. Product Code: 0042-0000-JP; GTIN: 813502010428; All Serial Numbers. 7. Product Code: 0042-0000-JP-L; GTIN: 813502010428; All Serial Numbers. 8. Product Code: 0042-0000-UK; GTIN: 813502011296; All Serial Numbers. 9. Product Code: 0042-0000-UK-L; GTIN: 813502011296; All Serial Numbers. 10. Product Code: 0042-0000-US; GTIN: 813502010022; All Serial Numbers. 11. Product Code: 0042-0000-US-L; GTIN: 813502010022; All Serial Numbers. 12. Product Code: 0042-0010-AU; GTIN: 813502010930; All Serial Numbers. 13. Product Code: 0042-0010-EU; GTIN: 813502010961; All Serial Numbers. 14. Product Code: 0042-0010-EU-L; GTIN: 813502010961; All Serial Numbers. 15. Product Code: 0042-0010-IN; GTIN: 813502011425; All Serial Numbers. 16. Product Code: 0042-0010-UK; GTIN: 813502010978; All Serial Numbers. 17. Product Code: 0042-0010-UK-L; GTIN: 813502010978; All Serial Numbers. 18. Product Code: 0042-0010-US; GTIN: 813502010985; All Serial Numbers. 19. Product Code: 0042-0010-US-L; GTIN: 813502010985; All Serial Numbers. 20. Product Code: 0042-0040-AU; GTIN: 813502012972; All Serial Numbers. 21. Product Code: 0042-0040-CA; GTIN: 813502012958; All Serial Numbers. 22. Product Code: 0042-0040-CT; GTIN: 813502011722; All Serial Numbers. 23. Product Code: 0042-0040-CT-L; GTIN: 813502011722; All Serial Numbers. 24. Product Code: 0042-0040-EU; GTIN: 813502012217; All Serial Numbers. 25. Product Code: 0042-0040-EU-L; GTIN: 813502012217; All Serial Numbers. 26. Product Code: 0042-0040-JP; GTIN: 813502011814; All Serial Numbers. 27. Product Code: 0042-0040-JP-L; GTIN: 813502011814; All Serial Numbers. 28. Product Code: 0042-0040-UK; GTIN: 813502012484; All Serial Numbers. 29. Product Code: 0042-0040-US; GTIN: 813502011401; All Serial Numbers. 30. Product Code: 0042-0040-US-L; GTIN: 813502011401; All Serial Numbers. 31. Product Code: 1000201; GTIN: 813502010442; All Serial Numbers. 32. Product Code: 1000432; GTIN: 813502013030; All Serial Numbers.
Nationwide distribution. International distribution to United Arab Emirates, Austria, Australia, Belgium, Brunei Darussalam, Brazil, Canada, Switzerland, Colombia, Czechia, Germany, Denmark, Spain, Finland, France, United Kingdom of Great Britain and Northern Ireland, Greece, Hong Kong, Croatia, Ireland, Israel, India, Italy, Japan, Kuwait, Luxembourg, Mexico, Malaysia, Netherlands, Norway, New Zealand, Panama, Pakistan, Poland, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Taiwan.
10,153 units