FDA Enforcement Class II Terminated

BD FACSLyric Flow Cytometer, 3-Laser 10 Color Instrument, REF: 662878, and 66287808 when using BD Trucount Tubes

Recall: Z-0069-2020 · Reported October 16, 2019

Enforcement

Recall Number
Z-0069-2020
Event ID
83603
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Becton, Dickinson and Company, BD Biosciences
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 16, 2019
Initiation Date
August 19, 2019
Classification Date
October 9, 2019
Termination Date
April 12, 2024
Address
2350 Qume Dr, N/A, San Jose, CA, 95131-1812, United States

Description

BD FACSLyric Flow Cytometer, 3-Laser 10 Color Instrument, REF: 662878, and 66287808 when using BD Trucount Tubes

Reason

Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.

Code Info

BD FACSLyric Flow Cytometer serial numbers: Z662878000003, Z662878000041, Z662878000042, Z662878000043, Z662878000044, Z662878000045, Z662878000046, Z662878000047, Z662878000049, Z662878000050, Z662878000051, Z662878000052, Z662878000053, Z662878000054, Z662878000055, Z662878000056, Z662878000057, Z662878000058, Z662878000059, Z662878000060, Z662878000061, Z662878000062, Z662878000063, Z662878000064, Z662878000065, Z662878000066, Z662878000067, Z662878000068, Z662878000069, Z662878000070, Z662878000071, Z662878000072, Z662878000073, Z662878000074, Z662878000075, Z662878000076, Z662878000077, Z662878000078, Z662878000079, Z662878000080, Z662878000081, Z662878000082, Z662878000083, Z662878000084, Z662878000085, Z662878000086, Z662878000087, Z662878000088, Z662878000089, Z662878000090, Z662878000091, Z662878000092, Z662878000093, Z662878000094, Z662878000095, Z662878000096, Z662878000097, Z662878000098, Z662878000105, Z662878000106, Z662878000107, Z662878000108, Z662878000109, Z662878000110, Z662878000111, Z662878000112, Z662878000113, Z662878000114, Z662878000115, Z662878000116, Z662878000117, Z662878000118, Z662878000119, Z662878000120, Z662878000121, Z662878000122, Z662878000123, Z662878000048.

Distribution

U.S.: CA, KS, NJ, MI, NY, PA, IL, IN, VA, WA, MD, IA, OK, RI, FL, CO, OH, TX, TN, GA, MO, MN, HI, VT, MA, NE. O.U.S. (Foreign): Germany, Italy, South Korea, Japan, United Kingdom, Poland, France, Belgium, Canada, Latvia, India, Taiwan, Ireland, Algeria, Israel, Iraq, Netherlands, Malaysia, Australia, Sweden, Colombia, Switzerland, Slovakia, Spain, Brazil, Austria, Peru, Iran, Croatia, Lithuania, Czech Republic, Saudi Arabia, Finland, Qatar, South Africa, Denmark, Turkey, Mexico, Indonesia, Ecuador, Vietnam, Singapore

Quantity

78