ICU Medical 14" (36 cm) Appx 3.6 ml, Bifuse Transfer Set w/Rotating Luer, REF: 14237-28 The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Enforcement
- Recall Number
- Z-0069-2019
- Event ID
- 80946
- Classification
- Class II
- Status
- Completed
- Product Type
- Devices
- Firm
- ICU Medical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 10, 2018
- Initiation Date
- August 20, 2018
- Classification Date
- October 4, 2018
- Address
- 951 Calle Amanecer, San Clemente, CA, 92673-6212, United States
Description
ICU Medical 14" (36 cm) Appx 3.6 ml, Bifuse Transfer Set w/Rotating Luer, REF: 14237-28 The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Design change implemented changing the device from a blood set with a filter to a transfer set without a filter. Both devices had the same listing number. Customers may not be aware of the product design change and use the transfer set as a blood transfusion set without realizing a blood filter is not included.
lots: 85-540-SJ, 86-189-SL, 86-911-SJ
U.S. Distribution in states of: NY, AL, CA, MS, PA, GA, MI, NC, WI, IL, WA, UT and FL.
2361 Units