VARISOURCE Breast Template System for high-dose rate Brachytherapy. ***INSTRUCTIONS FOR USE*** Breast Bridge & Template System AL13011000 ***Manufacturer: Varian Medical Systems, Inc., 3100 Hansen Way, Palo Alto, CA 94304 USA
Recall: Z-0065-2014
·
Reported October 30, 2013
Enforcement
- Recall Number
- Z-0065-2014
- Event ID
- 66121
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Varian Medical Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 30, 2013
- Initiation Date
- August 9, 2013
- Classification Date
- October 24, 2013
- Termination Date
- April 10, 2014
- Address
- 700 Harris St Ste 109, N/A, Charlottesville, VA, 22903-4584, United States
Description
VARISOURCE Breast Template System for high-dose rate Brachytherapy. ***INSTRUCTIONS FOR USE*** Breast Bridge & Template System AL13011000 ***Manufacturer: Varian Medical Systems, Inc., 3100 Hansen Way, Palo Alto, CA 94304 USA
Reason
The recall was initiated after Varian Medical became aware the Lexan templates of the Breast Bridge Template System could become bent and turn opaque after steam sterilization.
Code Info
Lots #6078184-1, #6078184-2, #6078184-4
Distribution
Worldwide distribution: US (nationwide) distribution in CA and in the country of India.
Quantity
3 sets