FDA Enforcement Class II Terminated

Symbia T Series SPECT + CT Product Usage 1) SPECT: To Detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588 keV. 2) CT: The CT component is intended to produce corss-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. 3) SPECT + CT: Perform CT scans and nuclear imaging studies with the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patients anatomy.

Recall: Z-0062-2017 · Reported October 19, 2016

Enforcement

Recall Number
Z-0062-2017
Event ID
66544
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Other
Report Date
October 19, 2016
Initiation Date
August 16, 2013
Classification Date
October 7, 2016
Termination Date
December 23, 2016
Address
2501 Barrington Rd, N/A, Hoffman Estates, IL, 60195-2061, United States

Description

Symbia T Series SPECT + CT Product Usage 1) SPECT: To Detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588 keV. 2) CT: The CT component is intended to produce corss-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. 3) SPECT + CT: Perform CT scans and nuclear imaging studies with the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patients anatomy.

Reason

The detectors on the Symbia S and Symbia T camera systems are supported and moved horizontally and vertically by four linear bearing cars. There is a potential for the linear bearing car to fail allowing the ball bearings to fall out.

Code Info

Material Number 10275007, Serial Numbers 1001 , 1006 , 1007 , 1032 , 1053 , 1104 , 1134 , 1012 , 1226 , 1014 , 1015 , 1284 , 1024 , 1034 , 1055 , 1060 , 1064 , 1038 , 1039 , 1065 , 1041 , 1071 , 1043 , 1044 , 1076 , 1098 , 1108 , 1109 , 1110 , 1050 , 1051 , 1111 , 1112 , 1113 , 1117 , 1056 , 1057 , 1119 , 1131 , 1137 , 1145 , 1146 , 1147 , 1160 , 1170 , 1066 , 1180 , 1181 , 1183 , 1189 , 1191 , 1199 , 1204 , 1074 , 1213 , 1215 , 1231 , 1232 , 1249 , 1252 , 1269 , 1287 , 1292 , 1318 , 1324 , 1086 , 1087 , 1325 , 1326 , 1090 , 1091 , 1327 , 1099 , 1101 , 1102 , 1116 , 1130 , 1042 , 1096 , 1097 , 1115 , 1173 , 1162 , 1235 , 1259 , 1309 , 1171 , 1175 , 1176 , 1178 , 1179 , 1185 , 1186 , 1036 , 1200 , 1202 , 1206 , 1209 , 1210 , 1211 , 1018 , 1026 , 1045 , 1216 , 1217 , 1058 , 1114 , 1128 , 1192 , 1193 , 1194 , 1195 , 1246 , 1257 , 1258 , 1291 , 1229 , 1307 , 1315 , 1234 , 1236 , 1242 , 1253 , 1262 , 1266 , 1267 , 1273 , 1276 , 1278 , 1282 , 1290 , 1295 , 1296 , 1310 , 1317 , 1319 , 1321 , 1322 , 1100 , 1208 , 1281 , 1062 , 1078 , 1009 , 1122 , 1132 , 1011 , 1017 , 1019 , 1077 , 1094 , 1106 , 1149 , 1168 , 1172 , 1221 , 1225 , 1251 , 1268 , 1305 , 1142 , 1254 , 1320 , 1247 , 1264 , 1285 , 1299 , 1125 , 1143 , 1190 , 1277 , 1005 , 1008 , 1021 , 1047 , 1049 , 1070 , 1073 , 1075 , 1105 , 1107 , 1124 , 1144 , 1150 , 1158 , 1163 , 1164 , 1165 , 1177 , 1182 , 1184 , 1188 , 1196 , 1197 , 1203 , 1214 , 1218 , 1219 , 1239 , 1244 , 1250 , 1263 , 1265 , 1271 , 1279 , 1288 , 1303 , 1306 , 1312 , 1313 , 1029 , 1030 , 1035 , 1046 , 1079 , 1085 , 1156 , 1159 , 1207 , 1224 , 1261 , 1272 , 1283 , 1151 , 1152 , 1153 , 1154 , 1169 , 1227 , 1311 , 1088 , 1135 , 1166 , 1238 , 1297 , 1010 , 1031 , 1120 , 1121 , 1123 , 1013 , 1061 , 1069 , 1083 , 1103 , 1167 , 1187 , 1223 , 1294 , 1016 , 1020 , 1025 , 1027 , 1054 , 1067 , 1068 , 1080 , 1081 , 1082 , 1089 , 1129 , 1161 , 1201 , 1205 , 1228 , 1230 , 1233 , 1240 , 1245 , 1286 , 1298 , 1300 , 1302 , 1304 , 1157 , 1040 , 1275 , 1023 , 1033 , 1092 , 1118 , 1139 , 1140 , 1141 , 1155 , 1212 , 1220 , 1222 , 1243 , 1255 , 1256 , 1270 , 1308 , 1260 , 1003 , 1028 , 1095 , 1241 , 1293 , 1314 , 1048 , 1063 , 1136 , 1289 , 1138 , 1198 , 1248 , 1148 , 1072 , 1127 , 1022 , 1323 , 1052 , 1084 , 1126 , 1301 , 1037 , 1059 , 1093 , 1133 , 1174 , 1237 , 1274 , 1316 , 46114 , 6053 , 68282 , 74401 , 74404 , 8473047245. Material Number 10275008, Serial Numbers 1071 , 1077 , 1051 , 1063 , 1124 , 1054 , 1172 , 1016 , 1032 , 1033 , 1060 , 1093 , 1171 , 1045 , 1049 , 1056 , 1082 , 1092 , 1095 , 1107 , 1111 , 1121 , 1123 , 1131 , 1138 , 1144 , 1145 , 1176 , 1182 , 1098 , 1002 , 1068 , 1166 , 1009 , 1035 , 1011 , 1037 , 1044 , 1059 , 1074 , 1083 , 1102 , 1103 , 1118 , 1141 , 1142 , 1149 , 1023 , 1156 , 1161 , 1170 , 1028 , 1029 , 1173 , 1178 , 1181 , 1186 , 1038 , 1040 , 1058 , 1090 , 1091 , 1094 , 1053 , 1097 , 1062 , 1099 , 1100 , 1104 , 1108 , 1113 , 1114 , 1117 , 1139 , 1150 , 1177 , 1165 , 1167 , 1168 , 1169 , 1179 , 1180 , 1184 , 1185 , 1566 , 1039 , 1042 , 1055 , 1065 , 1066 , 1096 , 1109 , 1127 , 1128 , 1153 , 1175 , 1003 , 1057 , 1080 , 1001 , 1004 , 1006 , 1007 , 1012 , 1013 , 1021 , 1024 , 1026 , 1036 , 1048 , 1052 , 1061 , 1070 , 1084 , 1087 , 1101 , 1105 , 1106 , 1135 , 1158 , 1010 , 1030 , 1031 , 1078 , 1143 , 1018 , 1020 , 1183 , 1120 , 1140 , 1017 , 1025 , 1064 , 1089 , 1188 , 1022 , 1047 , 1067 , 1075 , 1086 , 1116 , 1125 , 1126 , 1132 , 1136 , 1137 , 1151 , 1041 , 1085 , 1147 , 1008 , 1069 , 1115 , 1133 , 1160 , 1162 , 1164 , 1110 , 1073 , 1159 , 1088 , 1154 , 1076 , 1014 , 1015 , 1134 , 1072 , 1119 , 1122 , 1130 , 1146 , 1155 , 1174 , 1050 , 1079 , 1148 , 1081 , 1005 , 1157 , 1046 , 1019 , 1034 , 1129 , 1152 , 17283 , 17440 , 62322 , 7283 , 72836 , 72837 , 72862 , 7440 , 74409 , 7444 , 8473047245. Material Number 10275009, Serial Numbers 1140 , 1006 , 1044 , 1046 , 1089 , 1153 , 1195 , 1009 , 1014 , 1072 , 1090 , 1100 , 1132 , 1155 , 1178 , 1182 , 1041 , 1124 , 1024 , 1026 , 1069 , 1076 , 1091 , 1104 , 1120 , 1127 , 1145 , 1159 , 1123 , 1128 , 1033 , 1102 , 1106 , 1110 , 1156 , 1165 , 1187 , 1193 , 1194 , 1114 , 1086 , 1051 , 1080 , 1092 , 1149 , 1066 , 1001 , 1007 , 1008 , 1047 , 1060 , 1011 , 1013 , 1015 , 1017 , 1019 , 1022 , 1025 , 1027 , 1028 , 1029 , 1030 , 1034 , 1040 , 1093 , 1131 , 1133 , 1052 , 1056 , 1057 , 1064 , 1071 , 1077 , 1078 , 1082 , 1083 , 1084 , 1088 , 1094 , 1095 , 1098 , 1099 , 1101 , 1105 , 1108 , 1111 , 1036 , 1129 , 1079 , 1085 , 1103 , 1112 , 1134 , 1135 , 1115 , 1126 , 1139 , 1166 , 1168 , 1142 , 1170 , 1144 , 1173 , 1146 , 1179 , 1151 , 1152 , 1157 , 1158 , 1160 , 1161 , 1162 , 1164 , 1174 , 1180 , 1183 , 1184 , 1186 , 1190 , 1196 , 1012 , 1050 , 1068 , 1070 , 1109 , 1143 , 1004 , 1075 , 1031 , 1038 , 1065 , 1119 , 1167 , 1192 , 1113 , 1010 , 1016 , 1018 , 1020 , 1021 , 1023 , 1039 , 1042 , 1048 , 1063 , 1074 , 1081 , 1097 , 1117 , 1125 , 1148 , 1154 , 1163 , 1175 , 1191 , 1058 , 1141 , 1130 , 1172 , 1087 , 1107 , 1121 , 1122 , 1037 , 1053 , 1062 , 1137 , 1005 , 1043 , 1045 , 1061 , 1067 , 1138 , 1176 , 1032 , 1059 , 1150 , 1054 , 1116 , 1118 , 1147 , 1171 , 1185 , 1002 , 1055 , 1049 , 1177 , 1073 , 1181 , 1003 , 1035 , 1096 , 1169 , 1197 , 62323 , 7440 , 7502. Material Number 10275010, Serial Numbers 1161 , 1163 , 1059 , 1072 , 1006 , 1053 , 1061 , 1076 , 1078 , 1096 , 1108 , 1109 , 1113 , 1116 , 1117 , 1118 , 1119 , 1141 , 1142 , 1153 , 1160 , 1166 , 1173 , 1030 , 1049 , 1145 , 1150 , 1057 , 1058 , 1156 , 1012 , 1052 , 1149 , 1151 , 1005 , 1014 , 1079 , 1122 , 1168 , 1037 , 1045 , 1056 , 1066 , 1088 , 1121 , 1136 , 1155 , 1167 , 1125 , 1019 , 1022 , 1034 , 1044 , 1048 , 1050 , 1093 , 1101 , 1107 , 1112 , 1135 , 1064 , 1111 , 1120 , 1172 , 1103 , 1140 , 1152 , 1171 , 1001 , 1003 , 1004 , 1008 , 1009 , 1011 , 1013 , 1015 , 1016 , 1017 , 1018 , 1025 , 1027 , 1028 , 1031 , 1032 , 1038 , 1039 , 1042 , 1043 , 1051 , 1054 , 1055 , 1070 , 1071 , 1075 , 1002 , 1021 , 1026 , 1047 , 1085 , 1083 , 1084 , 1095 , 1086 , 1154 , 1089 , 1090 , 1094 , 1097 , 1098 , 1100 , 1102 , 1104 , 1105 , 1110 , 1115 , 1130 , 1137 , 1138 , 1139 , 1143 , 1164 , 1158 , 1170 , 1007 , 1024 , 1040 , 1065 , 1074 , 1082 , 1087 , 1091 , 1092 , 1133 , 1144 , 1147 , 1148 , 1162 , 1165 , 1174 , 1146 , 1035 , 1077 , 1134 , 1029 , 1036 , 1041 , 1063 , 1069 , 1080 , 1081 , 1114 , 1159 , 1010 , 1067 , 1124 , 1033 , 1129 , 1062 , 1073 , 1106 , 1126 , 1127 , 1128 , 1131 , 1132 , 1046 , 1020 , 1023 , 1157 , 1068 , 1060 , 1099 , 1123 , 1644 , 268045 , 72838 , 74401 . Material Number 8717733, Serial Numbers 1040 , 1088 , 1103 , 1157 , 1158 , 1159 , 1163 , 1041 , 1042 , 1043 , 1071 , 1072 , 1073 , 1074 , 1083 , 1090 , 1107 , 1147 , 1171 , 1173 , 1178 , 1196 , 1197 , 1009 , 1038 , 1066 , 1068 , 1070 , 1076 , 1079 , 1087 , 1128 , 1143 , 1144 , 1150 , 1153 , 1177 , 1188 , 1189 , 1193 , 1199 , 1108 , 1194 , 1000 , 1005 , 1057 , 1061 , 1094 , 1116 , 1141 , 1151 , 1183 , 1205 , 1025 , 1101 , 1024 , 1029 , 1044 , 1048 , 1053 , 1062 , 1081 , 1086 , 1106 , 1117 , 1135 , 1181 , 1209 , 701001 , 1058 , 1082 , 1105 , 1162 , 1064 , 1010 , 1015 , 1032 , 1050 , 1119 , 1013 , 1014 , 1016 , 1017 , 1027 , 1035 , 1036 , 1039 , 1051 , 1060 , 1078 , 1084 , 1085 , 1099 , 1109 , 1110 , 1111 , 1112 , 1122 , 1129 , 1132 , 1136 , 1140 , 1145 , 1155 , 1156 , 1164 , 1165 , 1166 , 1172 , 1186 , 1192 , 1195 , 1201 , 1120 , 1137 , 1146 , 1148 , 1149 , 1176 , 1185 , 1045 , 1095 , 1097 , 1004 , 1007 , 1008 , 1011 , 1019 , 1020 , 1021 , 1022 , 1023 , 1026 , 1028 , 1056 , 1059 , 1098 , 1034 , 1121 , 1200 , 1037 , 1052 , 1054 , 1055 , 1065 , 1067 , 1075 , 1077 , 1080 , 1091 , 1093 , 1113 , 1115 , 1118 , 1125 , 1126 , 1133 , 1139 , 1142 , 1089 , 1124 , 1130 , 1134 , 1160 , 1152 , 1154 , 1161 , 1167 , 1168 , 1169 , 1170 , 1175 , 1180 , 1182 , 1187 , 1191 , 1198 , 1202 , 1203 , 1204 , 1206 , 1208 , 1092 , 1012 , 1049 , 1096 , 1063 , 1033 , 1100 , 1174 , 1030 , 1127 , 1138 , 1184 , 1003 , 1104 , 1179 , 1190 , 1046 , 1001 , 1002 , 1006 , 1018 , 1031 , 1102 , 1114 , 1131 , 1210 , 733612 , D613 .

Distribution

Worldwide Distribution - US: Nationwide, including Washington D.C., and Puerto Rico. and the countries of: Poland, France, Austria, Czech Republic, Ireland, Germany, Italy, Japan, Canada, United Arab Emirates, Turkey, Netherlands, United Kingdom, Brazil, South Africa, Kuwait, Belgium, Russia, Taiwan, Australia, Bangladesh, Bulgaria, Brazil, Switzerland, Chile, China, Colombia, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Finland, Hong Kong, Croatia, Hungary, Indonesia, Iran, Israel, India, South Korea, Kuwait, Kazakhstan, Lithuania, Latvia, Morocco, Mexico, Norway, New Zealand, Peru, Phillipines, Pakistan, Portugal, Qatar, Romania, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Thailand, Tunisia, Taiwan, Venezuela, and Vietnam.

Quantity

1109 units