FDA Enforcement Class II Ongoing

VITROS Immunodiagnostic Products - TSH Reagent Pack (10758750000227) 1912997

Recall: Z-0061-2019 · Reported October 10, 2018

Enforcement

Recall Number
Z-0061-2019
Event ID
80717
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Ortho Clinical Diagnostics Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 10, 2018
Initiation Date
July 12, 2018
Classification Date
October 3, 2018
Address
1001 US Highway 202, Raritan, NJ, 08869-1424, United States

Description

VITROS Immunodiagnostic Products - TSH Reagent Pack (10758750000227) 1912997

Reason

The recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use (IFU). Patients who are taking biotin supplements could potentially have biased sample results for affected assays at biotin concentrations lower than indicated in the current IFU.

Code Info

5535 5540 5550 5560 5570 5580 5590 5600 5610 5615 5620 5630 5640 5650 5655 5660 5665 5670 5675 5678 5680 5685 5690

Distribution

Worldwide Distribution - US Nationwide distribution, and countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Singapore, United Kingdom, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, and Netherlands.

Quantity

68230