DJO EMPOWR Tibial Knee Punch Handle Part No. 801-05-040 and 801-05-018 contained in Instrument Tray for EMPOWR Porous Knee - Product Usage: The Tibial Punch Handle is used in conjunction with one of three appropriately sized Tibial Punch Guides to broach the tibial canal in preparation for the tibial baseplate keel.
Enforcement
- Recall Number
- Z-0060-2021
- Event ID
- 86257
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Encore Medical, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- October 14, 2020
- Initiation Date
- August 3, 2020
- Classification Date
- October 7, 2020
- Address
- 9800 Metric Blvd, N/A, Austin, TX, 78758-5445, United States
Description
DJO EMPOWR Tibial Knee Punch Handle Part No. 801-05-040 and 801-05-018 contained in Instrument Tray for EMPOWR Porous Knee - Product Usage: The Tibial Punch Handle is used in conjunction with one of three appropriately sized Tibial Punch Guides to broach the tibial canal in preparation for the tibial baseplate keel.
There is a potential for the retaining pin of the tibial punch handle to become loose and fall out while impacting. If undetected, there is a potential of leaving debris in the joint space.
Model/Catalog no. 801-05-040; :Lot/serial No.s 281905L08, 281905L15, 283808L01, 305197L02, 318871L10 Model/Catalog no. 801-05-018, Lot/serial No.s: 215488L08, 236760L01, 228231L01, 241946L02, 259710L01, 314969L01
U.S. Nationwide distribution including in the states of VA, TN, MN, TX, IL, IN, KY, OH, SC, RI, MD, PA, NY, CA, KS, OK, AR, LA, FL, AL, GA, AZ, NV, WA, MS, ME, NV and Puerto Rico.
190 units