Integra Licox Brain Tissue Oxygen Monitoring 1 x Complete Brain Probe Kit REF IM3.SR Rx Only, Manufacturer GMS- Gesellschaft fur medizinische, Sondentechnik mbH Dofstrasse 2 24247 Melkendorf, Germany Distributed by Integra Neurosciences 311 Enterprise Drive Plainsboro, NJ 08536 USA Used in direct monitoring of the partial pressure of oxygen.
Enforcement
- Recall Number
- Z-0058-2014
- Event ID
- 65742
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Integra LifeSciences Corp.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 30, 2013
- Initiation Date
- June 11, 2013
- Classification Date
- October 23, 2013
- Termination Date
- February 18, 2014
- Address
- 311 Enterprise Dr, N/A, Plainsboro, NJ, 08536-3344, United States
Description
Integra Licox Brain Tissue Oxygen Monitoring 1 x Complete Brain Probe Kit REF IM3.SR Rx Only, Manufacturer GMS- Gesellschaft fur medizinische, Sondentechnik mbH Dofstrasse 2 24247 Melkendorf, Germany Distributed by Integra Neurosciences 311 Enterprise Drive Plainsboro, NJ 08536 USA Used in direct monitoring of the partial pressure of oxygen.
Integra LifeSciences Corporation has determined that there is a possibility that some lots of Licox CMP Brain Tissue Oxygen and Temperature Catheter Kits contain a hex wrench that does not loosen the set screw on the drill bit.
K002765 Lot # 130712, 210912, 220612, 172281, 172647, 172690, 173170, 173260, 173548, 173580, 173917, 173942
Worldwide distribution: US (nationwide) and countries of: AR, BR, CH, CL,CR. DE, ES, FI, FR, GB, IL, IT, MX, PA, PL, PT, RU, TW, and ZA.
175 KITS