FDA Enforcement
Class II
Terminated
Green Mamba Suture Passer. The Biomet Sports Medicine Mamba instruments are utilized to aid in passing suture through soft tissue. The Mamba Needle is designed for use with the Mamba Suture Passer. The needles are sterile, disposable, and for single-patient use only.
Recall: Z-0057-2015
·
Reported October 15, 2014
Enforcement
- Recall Number
- Z-0057-2015
- Event ID
- 69360
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Biomet, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 15, 2014
- Initiation Date
- September 26, 2014
- Classification Date
- October 9, 2014
- Termination Date
- July 14, 2015
- Address
- 56 E Bell Dr, Warsaw, IN, 46581, United States
Description
Green Mamba Suture Passer. The Biomet Sports Medicine Mamba instruments are utilized to aid in passing suture through soft tissue. The Mamba Needle is designed for use with the Mamba Suture Passer. The needles are sterile, disposable, and for single-patient use only.
Reason
An investigation identified that high level friction may exist between the needle and Mamba suture passer instrument, causing the needle to break during use.
Code Info
Catalog Number: 110010850 Lot Number: 231120, 253210
Distribution
Nationwide Distribution-including the states of FL, NY, TX, VA, NV, MI, CA, IN, SD, GA, KY, NC, and AR.
Quantity
12 units