FDA Enforcement Class II Ongoing

Quick Release Gait Belt, 60

Recall: Z-0055-2021 · Reported October 14, 2020

Enforcement

Recall Number
Z-0055-2021
Event ID
86448
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Medline Industries Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
October 14, 2020
Initiation Date
September 15, 2020
Classification Date
October 5, 2020
Address
3 Lakes Dr, N/A, Northfield, IL, 60093-2753, United States

Description

Quick Release Gait Belt, 60

Reason

The buckle on the Gait Belt may break due to degradation of the material due to exposure to gamma radiation.

Code Info

Model: MDT821208NATM /503689 Lots: 33019040001, 33019090001 and 33019100001 Used in kits with model number: DYKF1045 Lots: 20EBB832 and 20EBM643.

Distribution

Domestic Distribution Only: IL, MI

Quantity

13,500 units