FDA Enforcement Class II Ongoing

Medline Sterile Laparoscopy Pack, DYNJ68187A (REF), containing the component ConMed/Unimax Detachable Endo Pocket

Recall: Z-0053-2025 · Reported October 16, 2024

Enforcement

Recall Number
Z-0053-2025
Event ID
95338
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
MEDLINE INDUSTRIES, LP - Northfield
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 16, 2024
Initiation Date
August 28, 2024
Classification Date
October 10, 2024
Address
3 Lakes Dr, Northfield, IL, 60093-2753, United States

Description

Medline Sterile Laparoscopy Pack, DYNJ68187A (REF), containing the component ConMed/Unimax Detachable Endo Pocket

Reason

CONMED Corporation (vendor) initiated a recall for select lot numbers of the Unimax Detachable Endo Pocket. The Detachable Endo Pocket is indicated for use as a receptacle for the collection and extraction of tissues, organs, and calculi during general and laparoscopic procedures. The tube inside the pouch, which detaches during the removal process, may stretch out from the pouch opening if it is not precisely fixed during the production process. Medline has identified that this product is utilized as a component within the Medline-branded convenience kits.

Code Info

Lot numbers 24AMA209, exp. 2025-06-30; 24BMB025, exp. 2025-08-31; 23LMA745, exp. 2025-07-31; UDI-DI each 10195327384494, UDI-DI case-40195327384495

Distribution

US Nationwide distribution in the state of CA.

Quantity