Monarch Platform (Urology), REF: MON-000007, MON-000007-RFB
Enforcement
- Recall Number
- Z-0052-2025
- Event ID
- 95336
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- AURIS HEALTH INC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 16, 2024
- Initiation Date
- August 2, 2024
- Classification Date
- October 10, 2024
- Address
- 150 Shoreline Dr, Redwood City, CA, 94065-1400, United States
Description
Monarch Platform (Urology), REF: MON-000007, MON-000007-RFB
Endoscopy/bronchoscopy/urology devices may have a strain relief defect with some cart and tower power cords that may cause an electrical short, which may lead to system shutdown, which can lead to prolonged procedure or case conversion, and if the user (e.g., operating room staff) touches exposed wires or conductive fluids in contact with the exposed wires, there is a chance of electrical shock.
Cart Power Cord 20' 14AWG BLK, PN 304-003613-00. Tower Power Cord 20' 14AWG BLK, PN 304-003607-00. REF/UDI-DI/Serial Numbers: MON-000007/+B634MON0000070/120059, 120133, 210002; MON-000007-RFB/10810068810889/210001.
Worldwide - US Nationwide distribution including in the states of MN, FL, NC, AK, NJ, NY, VA, WI, AZ, MS, LA, MI, MA, MT, WV, GA, DE, TX, CA, OH, IN, SC, PA, CT, ND, MO, DC, IL, NE, ID, WA, TN, OR, AL, SD, CO, KY, NM, VT, MD and the countries of Canada.
4