FDA Enforcement
Class I
Ongoing
GE Heathcare Giraffe OmniBed; neo natal incubator
Recall: Z-0047-2025
·
Reported October 23, 2024
Enforcement
- Recall Number
- Z-0047-2025
- Event ID
- 95282
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- DATEX--OHMEDA, INC.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 23, 2024
- Initiation Date
- September 3, 2024
- Classification Date
- October 17, 2024
- Address
- 9900 W Innovation Dr, Wauwatosa, WI, 53226-4856, United States
Description
GE Heathcare Giraffe OmniBed; neo natal incubator
Reason
GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.
Code Info
All models
Distribution
Worldwide distribution.