FDA Enforcement Class II Terminated

STAR ActiveTrak Excimer Laser System, Model: STAR S3, Catalog Number: 0030-2450, Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA The STAR Excimer Laser Systems are indicated for laser assisted in situ keratomileusis (LASIK), photorefractive keratectomy (PRK), and phototherapeutic keratectomy (PTK) eye surgery in certain patient populations.

Recall: Z-0047-2014 · Reported October 30, 2013

Enforcement

Recall Number
Z-0047-2014
Event ID
66304
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Abbott Medical Optics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 30, 2013
Initiation Date
September 16, 2013
Classification Date
October 22, 2013
Termination Date
November 12, 2014
Address
510 Cottonwood Dr, N/A, Milpitas, CA, 95035-7403, United States

Description

STAR ActiveTrak Excimer Laser System, Model: STAR S3, Catalog Number: 0030-2450, Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA The STAR Excimer Laser Systems are indicated for laser assisted in situ keratomileusis (LASIK), photorefractive keratectomy (PRK), and phototherapeutic keratectomy (PTK) eye surgery in certain patient populations.

Reason

This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the AC Power Distribution Printed Circuit Board (PCB).

Code Info

All units of these models: STAR S2 Excimer Laser System; STAR S3 ActiveTrak; STAR S4 with Variable Spot Scanning; and STAR S41R; There are no expiration dates for the STAR Excimer Laser Systems.

Distribution

Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Asia Pacific, Canada, Europe, Africa, Middle East, Japan, and Latin America.

Quantity

834 units in total