FDA Enforcement Class II Terminated

Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation.

Recall: Z-0046-2018 · Reported November 1, 2017

Enforcement

Recall Number
Z-0046-2018
Event ID
77982
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Beckman Coulter Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Other
Report Date
November 1, 2017
Initiation Date
August 3, 2017
Classification Date
October 24, 2017
Termination Date
September 27, 2023
Address
250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States

Description

Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation.

Reason

Beckman Coulter has identified that due to a PLC software nonconformity the 1K Stockyard can initiate a retrieval of a sample tube during the rack loading process, which should not occur. This issue and associated complaint were discovered and filed internally.

Code Info

Software version PLC V26R10

Distribution

China France Italy Kuwait Spain Turkey United States

Quantity

27 units