FDA Enforcement
Class II
Terminated
Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation.
Recall: Z-0046-2018
·
Reported November 1, 2017
Enforcement
- Recall Number
- Z-0046-2018
- Event ID
- 77982
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Beckman Coulter Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Other
- Report Date
- November 1, 2017
- Initiation Date
- August 3, 2017
- Classification Date
- October 24, 2017
- Termination Date
- September 27, 2023
- Address
- 250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States
Description
Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation.
Reason
Beckman Coulter has identified that due to a PLC software nonconformity the 1K Stockyard can initiate a retrieval of a sample tube during the rack loading process, which should not occur. This issue and associated complaint were discovered and filed internally.
Code Info
Software version PLC V26R10
Distribution
China France Italy Kuwait Spain Turkey United States
Quantity
27 units