Sysmex Fluorocell(R) PLT Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzers
Enforcement
- Recall Number
- Z-0045-2018
- Event ID
- 78134
- Classification
- Class II
- Status
- Completed
- Product Type
- Devices
- Firm
- Sysmex America, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 1, 2017
- Initiation Date
- September 8, 2017
- Classification Date
- October 24, 2017
- Address
- 577 Aptakisic Rd, N/A, Lincolnshire, IL, 60069-4325, United States
Description
Sysmex Fluorocell(R) PLT Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzers
There is potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%, #) results obtained using the lots listed above may exhibit lower than actual PLT-F and IPF values. The issue may be recognized by a large discrepancy between the impedance platelet (PLT-I) result and the PLT-F result. In most cases, the falsely decreased PLT-F results display a PLT Abn Scattergram message with an asterisk beside the result or an Action Message: Difference between PLT and PLT- F. Check the results. These are indicators of unreliable data. In rare instances, platelet flags may be absent. The PLT-I results are unaffected.
Lot Numbers: A6051 A6052 A6053 A6054 A6055 A6056 A6057 A6058 A6059 A6060 A6061 A6062 A6063 A6064 A6065 A6066 A6067 A6068 A6069 A6070 A6071 A6072 A6073 A6074 A6075 A6076 A6077 A6078 A6079 A6080 A6081 A6082 A6083 A6084 A6085 A6086 A6087 A6088 A6089 A6090
Worldwide Distribution - US (nationwide including Puerto Rico) Internationally to New Zealand, Brazil, India, Philippines, Nicaragua, Argentina, Australia, Chile, Viet Nam, Colombia, Pakistan, Malaysia, Dubai, Germany, Hong Kong, Bangladesh, Uruguay Thailand, China, Saudi Arabia, Taiwan, Honduras, Ecuador, Kuwait, South Korea, and Costa Rica
23,239 units (9353 US)