FDA Enforcement
Class II
Terminated
STERRAD 100NX Cassettes The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.
Recall: Z-0044-2014
·
Reported October 30, 2013
Enforcement
- Recall Number
- Z-0044-2014
- Event ID
- 66317
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Advanced Sterilization Products
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 30, 2013
- Initiation Date
- September 18, 2013
- Classification Date
- October 22, 2013
- Termination Date
- January 13, 2014
- Address
- 33 Technology Dr, N/A, Irvine, CA, 92618-2346, United States
Description
STERRAD 100NX Cassettes The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.
Reason
Advanced Sterilization Products (ASP) is recalling the three lots (13F049, 13F050, 13F051) STERRAD 100NX Cassettes because ASP has determined that the leak indicator on the cassettes is effective for only 12 months of the cassette's labeled 15-month shelf-life.
Code Info
Lot# 13F050, 13F051, 13F049
Distribution
Nationwide Distribution
Quantity
2,995 units