NxTAG Respiratory Pathogen Panel +SARS-CoV-2, REF: I056C0468
Recall: Z-0043-2025
·
Reported October 16, 2024
Enforcement
- Recall Number
- Z-0043-2025
- Event ID
- 95377
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Luminex Molecular Diagnostics Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 16, 2024
- Initiation Date
- September 5, 2024
- Classification Date
- October 9, 2024
- Address
- 439 University Ave Suite 900, Toronto, N/A, Canada
Description
NxTAG Respiratory Pathogen Panel +SARS-CoV-2, REF: I056C0468
Reason
Respiratory pathogen panel may report an elevated rate (7.6% predicted probability of occurrence) of non-specific hMPV target (human Metapneumovirus) results, risk identified predominantly in clinical samples that present positivity for another target, leading to an erroneous co-infection result, can potentially result in hMPV false positives.
Code Info
Lot/Expiration: IK056C-0033/ 9/30/2024, IK056C-0034/ 10/31/2024, IK056C-0035/ 10/31/2024, IK056C-0036/ 12/31/2024, IK056C-0037/ 12/31/2024, IK056C-0038/ 12/31/2024, IK056C-0039/ 1/31/2025
Distribution
U Nationwide distribution in the states of OH, CA, NC, WI, TN, TX, MI, NY, MO, FL, MA, NJ, PA, VA, AZ, IL, OK, CO, MN.
Quantity
613