FDA Enforcement Class II Ongoing

NxTAG Respiratory Pathogen Panel, REF: I051C0447

Recall: Z-0042-2025 · Reported October 16, 2024

Enforcement

Recall Number
Z-0042-2025
Event ID
95377
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Luminex Molecular Diagnostics Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 16, 2024
Initiation Date
September 5, 2024
Classification Date
October 9, 2024
Address
439 University Ave Suite 900, Toronto, N/A, Canada

Description

NxTAG Respiratory Pathogen Panel, REF: I051C0447

Reason

Respiratory pathogen panel may report an elevated rate (7.6% predicted probability of occurrence) of non-specific hMPV target (human Metapneumovirus) results, risk identified predominantly in clinical samples that present positivity for another target, leading to an erroneous co-infection result, can potentially result in hMPV false positives.

Code Info

UDI-DI: 00840487100417. Lot/Expiration: IK051C-0181/ 9/30/2024, IK051C-0182/ 9/30/2024, IK051C-0185/ 9/30/2024, IK051C-0186/ 9/30/2024, IK051C-0187/ 10/31/2024, IK051C-0188/ 10/31/2024, IK051C-0189/ 10/31/2024, IK051C-0190/ 10/31/2024, IK051C-0191/ 10/31/2024, IK051C-0192/ 10/31/2024, IK051C-0193/ 10/31/2024, IK051C-0194/ 10/31/2024, IK051C-0195/ 10/31/2024, IK051C-0196/ 11/30/2024, IK051C-0197/ 12/31/2024

Distribution

U Nationwide distribution in the states of OH, CA, NC, WI, TN, TX, MI, NY, MO, FL, MA, NJ, PA, VA, AZ, IL, OK, CO, MN.

Quantity

3375