FDA Enforcement
Class II
Ongoing
Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299
Recall: Z-0041-2024
·
Reported October 11, 2023
Enforcement
- Recall Number
- Z-0041-2024
- Event ID
- 93046
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Blue Belt Technologies, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 11, 2023
- Initiation Date
- September 7, 2023
- Classification Date
- October 5, 2023
- Address
- 2875 Railroad St, N/A, Pittsburgh, PA, 15201, United States
Description
Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299
Reason
CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon chooses to define the femoral rotational reference via the posterior condylar axis
Code Info
UDI: (01) 00885556773338 Lot Numbers: 51141585, 51108475, 51108689
Distribution
US Nationwide distribution.
Quantity