Citrine Triple Quad Mass Spectrometer, Part Number 5063684
Enforcement
- Recall Number
- Z-0040-2024
- Event ID
- 92810
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- AB Sciex, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 11, 2023
- Initiation Date
- August 30, 2023
- Classification Date
- October 5, 2023
- Address
- 500 Old Connecticut Path, N/A, Framingham, MA, 01701-4574, United States
Description
Citrine Triple Quad Mass Spectrometer, Part Number 5063684
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
UDI-DI: 00842365100533; Serial Numbers: DE220001803PL, DE230011804, DE230031806, DE230041807, DE230051808, DE230071901, DE230081903, DE230091903, DE230162003, DE230272105, DE230292105, DE240002106, DE240022111, DE240042203
US Nationwide.