FDA Enforcement Class II Ongoing

Citrine Triple Quad Mass Spectrometer, Part Number 5063684

Recall: Z-0040-2024 · Reported October 11, 2023

Enforcement

Recall Number
Z-0040-2024
Event ID
92810
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
AB Sciex, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 11, 2023
Initiation Date
August 30, 2023
Classification Date
October 5, 2023
Address
500 Old Connecticut Path, N/A, Framingham, MA, 01701-4574, United States

Description

Citrine Triple Quad Mass Spectrometer, Part Number 5063684

Reason

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

Code Info

UDI-DI: 00842365100533; Serial Numbers: DE220001803PL, DE230011804, DE230031806, DE230041807, DE230051808, DE230071901, DE230081903, DE230091903, DE230162003, DE230272105, DE230292105, DE240002106, DE240022111, DE240042203

Distribution

US Nationwide.

Quantity

14