FDA Enforcement Class II Ongoing

Citrine QTRAP Mass Spectrometer, Part Number 5063685

Recall: Z-0039-2024 · Reported October 11, 2023

Enforcement

Recall Number
Z-0039-2024
Event ID
92810
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
AB Sciex, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 11, 2023
Initiation Date
August 30, 2023
Classification Date
October 5, 2023
Address
500 Old Connecticut Path, N/A, Framingham, MA, 01701-4574, United States

Description

Citrine QTRAP Mass Spectrometer, Part Number 5063685

Reason

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

Code Info

UDI-DI: 00842365100540; Serial Numbers: DD250051901, DD250172009, DD250182009, DD230001804, DD250111910, DD250132004, DD250091909, DD250101910, DD250192010, CN10011707PT

Distribution

US Nationwide.

Quantity

10