FDA Enforcement
Class II
Completed
Pilling Knife Handles: a) REF 352950 b) REF 352953 c) REF 352951 d) REF 352957 e) REF 352952 : Reusable surgical knife handles, which are intended for use with blades which are inserted into a slot in the handle.
Recall: Z-0039-2018
·
Reported October 25, 2017
Enforcement
- Recall Number
- Z-0039-2018
- Event ID
- 78207
- Classification
- Class II
- Status
- Completed
- Product Type
- Devices
- Firm
- Teleflex Medical
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 25, 2017
- Initiation Date
- September 26, 2014
- Classification Date
- October 19, 2017
- Address
- 2917 Weck Dr, Research Triangle Park, NC, 27709-0186, United States
Description
Pilling Knife Handles: a) REF 352950 b) REF 352953 c) REF 352951 d) REF 352957 e) REF 352952 : Reusable surgical knife handles, which are intended for use with blades which are inserted into a slot in the handle.
Reason
The knife handles slot depth is out of specification, so blades may not fit on the handles properly.
Code Info
a) REF 352950: Lot numbers: D6, J6, E6, F6, G6, H6, I6 b) REF 352953: Lot numbers: I6, B6, E6 c) REF 352951: Lot numbers: I6, J6 d) REF 352957: Lot numbers: A6, B6, B7, K6, L6 e) REF 352952: Lot numbers: H6
Distribution
Worldwide Distribution: US (Nationwide) and countries of: Canada, France, Taiwan, Argentina, Australia, S. Korea, Belgium, Singapore, India, Indonesia, Philippines and Thailand.
Quantity
6197 units