FDA Enforcement Class II Terminated

Xoran Minicat00158

Recall: Z-0039-2014 · Reported November 6, 2013

Enforcement

Recall Number
Z-0039-2014
Event ID
66338
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Xoran Technologies, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 6, 2013
Initiation Date
July 27, 2010
Classification Date
October 25, 2013
Termination Date
September 11, 2014
Address
5210 S State Rd, N/A, Ann Arbor, MI, 48108-7936, United States

Description

Xoran Minicat00158

Reason

The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evaluation of the hazards reasonably related to defect or the failure to comply with the Federal standard; An evaluation of the health ha

Code Info

minicat

Distribution

Worldwide Distribution - US nationwide

Quantity

N/A