FDA Enforcement
Class II
Ongoing
ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile Ultrasound, ImaCor model ZHH.
Recall: Z-0038-2026
·
Reported October 15, 2025
Enforcement
- Recall Number
- Z-0038-2026
- Event ID
- 97505
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- ImaCor Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- October 15, 2025
- Initiation Date
- August 29, 2025
- Classification Date
- October 3, 2025
- Address
- 50 Jericho Tpke Ste 105, Jericho, NY, 11753-1014, United States
Description
ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile Ultrasound, ImaCor model ZHH.
Reason
Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.
Code Info
Model CLT-010; UDI-DI 00861589000108; Lot number: 20261209.
Distribution
US Nationwide distribution in the states of South Carolina, New Jersey, Pennsylvania, Florida, and Georgia.
Quantity
45 units