4500MD Triple Quad Mass Spectrometer, Part Number 5032522
Enforcement
- Recall Number
- Z-0038-2024
- Event ID
- 92810
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- AB Sciex, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 11, 2023
- Initiation Date
- August 30, 2023
- Classification Date
- October 5, 2023
- Address
- 500 Old Connecticut Path, N/A, Framingham, MA, 01701-4574, United States
Description
4500MD Triple Quad Mass Spectrometer, Part Number 5032522
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
UDI-DI: 00842365102049; Serial Numbers: EJ230051806PL, EJ230061807PL, EJ230071808, EJ230101810, EJ230111811, EJ230121811, EJ230141811, EJ230151812, EJ230161812, EJ230171901, EJ230181901, EJ230191901, EJ230211905, EJ230221908, EJ230231908, EJ230251909, EJ230271910, EJ230281910, EJ230311911, EJ230331912, EJ230362003, EJ230372003, EJ230392004, EJ230412009, EJ230422011, EJ230442012, EJ230452012, EJ230472101, EJ230482103, EJ230512105, EJ240012110, EJ240022111, EJ240052111, EJ240082207, EJ240092209, EJ240142307, EJ240132305, EJ240062201