FDA Enforcement
Class II
Terminated
Xoran Minicat00202
Recall: Z-0038-2014
·
Reported November 6, 2013
Enforcement
- Recall Number
- Z-0038-2014
- Event ID
- 66338
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Xoran Technologies, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 6, 2013
- Initiation Date
- July 27, 2010
- Classification Date
- October 25, 2013
- Termination Date
- September 11, 2014
- Address
- 5210 S State Rd, N/A, Ann Arbor, MI, 48108-7936, United States
Description
Xoran Minicat00202
Reason
The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evaluation of the hazards reasonably related to defect or the failure to comply with the Federal standard; An evaluation of the health ha
Code Info
minicat
Distribution
Worldwide Distribution - US nationwide
Quantity
320