FDA Enforcement Class II Terminated

PREMIER(R) Anterior Cervical Plate System SCREW BLOCK, REF 1860503, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132. Orthopedic. This instrument is a precision device which may incorporate a measuring function and has uses as described on the label. Unless labeled for single use, this instrument may be re-used.

Recall: Z-0038-2013 · Reported October 17, 2012

Enforcement

Recall Number
Z-0038-2013
Event ID
63157
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Medtronic Sofamor Danek USA Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
October 17, 2012
Initiation Date
August 30, 2012
Classification Date
October 10, 2012
Termination Date
August 15, 2013
Address
1800 Pyramid Place, N/A, Memphis, TN, 38132, United States

Description

PREMIER(R) Anterior Cervical Plate System SCREW BLOCK, REF 1860503, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132. Orthopedic. This instrument is a precision device which may incorporate a measuring function and has uses as described on the label. Unless labeled for single use, this instrument may be re-used.

Reason

Lines on the screw gauge may not placed on the caddy properly which could result in improper measurement of screws. An incorrect gauge could measure a screw to be approximately 1 mm shorter than its actual length.

Code Info

All lots except those starting with SY

Distribution

Worldwide Distribution-USA (nationwide) and the countries of India, Netherlands, Hong Kong, S. L Korea, Singapore, Canada, Australia, Japan, and Mexico.

Quantity

247 units