FDA Enforcement
Class II
Ongoing
API3200MD Mass Spectrometer, Part Number 4466230
Recall: Z-0037-2024
·
Reported October 11, 2023
Enforcement
- Recall Number
- Z-0037-2024
- Event ID
- 92810
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- AB Sciex, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 11, 2023
- Initiation Date
- August 30, 2023
- Classification Date
- October 5, 2023
- Address
- 500 Old Connecticut Path, N/A, Framingham, MA, 01701-4574, United States
Description
API3200MD Mass Spectrometer, Part Number 4466230
Reason
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
Code Info
Serial Numbers: EQ230011907, EQ230021911, EQ230001901
Distribution
US Nationwide.
Quantity
3