FDA Enforcement
Class III
Terminated
FSE Image Upgrade Kit, Catalog # 92019C This kit is used to install and/or upgrade software used with the EVOLIS Microplate System.
Recall: Z-0032-2014
·
Reported October 30, 2013
Enforcement
- Recall Number
- Z-0032-2014
- Event ID
- 66293
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Bio-Rad Laboratories
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 30, 2013
- Initiation Date
- August 20, 2013
- Classification Date
- October 21, 2013
- Termination Date
- November 13, 2013
- Address
- 6565 185th Ave NE, N/A, Redmond, WA, 98052-5039, United States
Description
FSE Image Upgrade Kit, Catalog # 92019C This kit is used to install and/or upgrade software used with the EVOLIS Microplate System.
Reason
The FSE Image Upgrade kit (part 92019C) is recalled due to issue related to the export of data from the EVOLIS analyzer to Laboratory Information Systems (LIS) when using the MONOLISA Anti-HBs EIA Quantitative Determination in EVOLIS APF version 6.0.
Code Info
Lot 122822
Distribution
US Distribution including the states of FL, IA, ME, MO, MS, NE, NY, TN, TX, and WA
Quantity
23 kits