FDA Enforcement Class I Ongoing

NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

Recall: Z-0029-2026 · Reported October 15, 2025

Enforcement

Recall Number
Z-0029-2026
Event ID
97516
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
NOXBOX LTD
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 15, 2025
Initiation Date
September 9, 2025
Classification Date
October 6, 2025
Address
Unit 1, Eurolink Gateway; Castle Road, Sittingbourne, N/A, United Kingdom

Description

NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

Reason

The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.

Code Info

Model Numbers: (1) NOXBOXI (NBL (Manufacturer)), (2) REQNOXBOXI (Linde (U.S. Distributor)); UDI-DI: 05060541640009; All serial numbers.

Distribution

Domestic: TN;

Quantity

1667 units