FDA Enforcement Class II Terminated

Universal Compact Head Ring Adapter Plate (UCHRAP), a component of the Universal Compact Head Ring Adapter, UCHRA. The UCHRA is a component of the Integra¿ CRW¿ Precision Arc Stereotactic System.

Recall: Z-0025-2016 · Reported October 14, 2015

Enforcement

Recall Number
Z-0025-2016
Event ID
72068
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Integra LifeSciences Corp.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 14, 2015
Initiation Date
August 21, 2015
Classification Date
October 6, 2015
Termination Date
October 13, 2016
Address
311 Enterprise Dr, N/A, Plainsboro, NJ, 08536-3344, United States

Description

Universal Compact Head Ring Adapter Plate (UCHRAP), a component of the Universal Compact Head Ring Adapter, UCHRA. The UCHRA is a component of the Integra¿ CRW¿ Precision Arc Stereotactic System.

Reason

Integra identified that the UCHRAP component (Arc Adapter Plate) would not assemble properly to the UCHRAR component (ARC Adapter Ring) of the Compact Head Ring Adapter due to an error in the manufacturing drawing.

Code Info

Cat. No. UCHRAP; Serial No.s 900, 901, 902, 904, 905, 906

Distribution

Worldwide Distribution - US Distribution to the states of : NJ, TX and WY., and to the countries of : Algeria, Germany and India.

Quantity

6 units