Proclaim DRG Implantable Pulse Generator Model 3664
Enforcement
- Recall Number
- Z-0024-2018
- Event ID
- 78154
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- St. Jude Medical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 25, 2017
- Initiation Date
- September 12, 2017
- Classification Date
- October 17, 2017
- Termination Date
- April 16, 2020
- Address
- 6901 Preston Rd, Plano, TX, 75024-2508, United States
Description
Proclaim DRG Implantable Pulse Generator Model 3664
Analysis of the difference between the actual device longevity and the battery indicator status provided by the Clinician Programmer (CP) or Patient Controller (PC) has revealed errors in the CP or PC longevity estimations that can overestimate or under estimate the battery indicator status in certain situations.
Serials implanted prior to September 12, 2017 are considered part of this advisory; Serial Numbers - AAP114.1 AAP267.1 AAP380.1 AAT623.1 ABM656.1 ABT952.1 ACD631.1 ACE905.1 ADJ588.1 ADJ614.1 ADJ842.1 ADJ852.1 ADJ855.1 ADJ863.1 ADJ872.1 ADJ910.1 ADK057.1 ADK078.1 ADK116.1 ADK186.1 ADK293.1 ADN214.1 ADN311.1 ADN340.1 ADN347.1 ADN376.1 ADN398.1 ADR256.1 ADR258.1 ADR260.1 ADR262.1 ADR289.1 ADR299.1 ADR302.1 ADR313.1 ADR362.1 ADR364.1 ADR365.1 ADR400.1 ADR401.1 ADR405.1 ADR410.1 ADR416.1 ADR421.1 ADR433.1 ADR434.1 ADR453.1 ADR457.1 ADR460.1 ADR461.1 ADR463.1 ADR464.1 ADR466.1 ADR469.1 ADR474.1 ADR475.1 ADR477.1 ADR485.1 ADR489.1 ADR492.1 ADR495.1 ADR497.1 ADR498.1 ADR501.1 ADR505.1 ADR509.1 ADR517.1 ADR519.1 ADR526.1 ADR543.1 ADR552.1 ADR557.1 ADR558.1 ADR563.1 ADR566.1 ADR575.1 ADR581.1 ADR598.1 ADR617.1 ADR630.1 ADR632.1 ADR655.1 ADR975.1 ADR986.1 ADS012.1 ADS039.1 ADW504.1 ADW624.1 ADW635.1 ADW686.1 ADW706.1 ADW707.1 ADW724.1 ADW930.1 AEH786.1 AEH900.1 AEJ018.1 AEJ059.1 AEM902.1 AFB279.1 AFB456.1 AFB710.1 AFB717.1 AFB934.1 AFG268.1 AFG397.1 AFG415.1 AFG464.1 AFG577.1 AFG587.1 AFG606.1 AFG780.1 AFG805.1 AFH511.1 AFH513.1 AFH514.1 AFH516.1 AFH518.1 AFH519.1 AFH521.1 AFH522.1 AFH524.1 AFH526.1 AFH527.1 AFH528.1 AFH530.1 AFH535.1 AFH538.1 AFH539.1 AFH543.1 AFH544.1 AFH545.1 AFH546.1 AFH547.1 AFH552.1 AFH554.1 AFH555.1 AFH556.1 AFH557.1 AFH558.1 AFH560.1 AFH561.1 AFH562.1 AFH563.1 AFH564.1 AFH565.1 AFH566.1 AFH570.1 AFH571.1 AFH572.1 AFH576.1 AFH579.1 AFH580.1 AFH582.1 AFH583.1 AFH584.1 AFH588.1 AFH589.1 AFH590.1 AFH592.1 AFH593.1 AFH594.1 AFH596.1 AFH597.1 AFH598.1 AFH599.1 AFH600.1 AFH601.1 AFH602.1 AFH605.1 AFH606.1 AFH607.1 AFH608.1 AFH609.1 AFH610.1 AFH614.1 AFH615.1 AFH618.1 AFH619.1 AFH620.1 AFH621.1 AFH624.1 AFH625.1 AFH626.1 AFH628.1 AFH629.1 AFH630.1 AFH631.1 AFH632.1 AFH633.1 AFH636.1 AFH641.1 AFH642.1 AFH646.1 AFH647.1 AFH648.1 AFH651.1 AFH654.1 AFH660.1 AFH663.1 AFH664.1 AFH665.1 AFH666.1 AFH668.1 AFH669.1 AFH671.1 AFH673.1 AFH674.1 AFH676.1 AFH684.1 AFH685.1 AFH687.1 AFH689.1 AFH695.1 AFH696.1 AFH697.1 AFH699.1 AFH700.1 AFH701.1 AFH702.1 AFH703.1 AFH706.1 AFH707.1 AFH708.1 AFZ281.1 AFZ365.1 AFZ581.1 AFZ582.1 AFZ592.1 AFZ620.1 AFZ631.1 AFZ637.1 AFZ642.1 AFZ649.1 AFZ661.1 AFZ685.1 AFZ691.1 AFZ695.1 AFZ710.1 AFZ711.1 AFZ728.1 AFZ730.1 AFZ768.1 AFZ780.1 AFZ870.1 AFZ881.1 AFZ893.1 AFZ908.1 AFZ915.1 AFZ918.1 AFZ930.1 AFZ933.1 AFZ970.1 AFZ972.1 AFZ978.1 AFZ979.1 AFZ981.1 AFZ998.1 AGA010.1 AGA012.1 AGA015.1 AGA029.1 AGA031.1 AGA032.1 AGA040.1 AGA044.1 AGA047.1 AGA049.1 AGA053.1 AGA061.1 AGA064.1 AGA067.1 AGA068.1 AGA071.1 AGA078.1 AGA083.1 AGA085.1 AGA087.1 AGA088.1 AGA094.1 AGA104.1 AGA108.1 AGA110.1 AGA113.1 AGM588.1 AGM603.1 AGN178.1 AGN273.1 AGN393.1 AGN394.1 AGN429.1 AGN476.1 AGN514.1 AGU357.1 AHD592.1 AHD733.1 AHD737.1 AHD739.1 AHD742.1 AHD745.1 AHD756.1 AHD759.1 AHD760.1 AHD763.1 AHD766.1 AHD775.1 AHD787.1 AHD795.1 AHD800.1 AHK478.1 AHK479.1 AHK480.1 AHK483.1 AHK485.1 AHK489.1 AHK491.1 AHK493.1 AHK498.1 AHK507.1 AHK516.1 AHK520.1 AHK527.1 AHK529.1 AHK533.1 AHK538.1 AHK540.1 AHK549.1 AHK570.1 AHK574.1 AHK583.1 AHK591.1 AHK601.1 AHK605.1 AHK610.1 AHK615.1 AHK655.1 AHK656.1 AHK688.1 AHK689.1 AHK722.1 AHK733.1 AHK736.1 AHK743.1 AHK763.1 AHK774.1 AHK812.1 AHK826.1 AHK959.1 AHL025.1 AHL058.1 AHL172.1 AHL182.1 AHV709.1 AHV710.1 AHV714.1 AHV715.1 AHV720.1 AHV732.1 AHV742.1 AHV752.1 AHV754.1 AHV757.1 AHV759.1 AHV761.1 AHV765.1 AHV766.1 AHV770.1 AHV777.1 AHV781.1 AHV995.1 AHW119.1 AHW177.1 AHW181.1 AHW187.1 AHW189.1 AHW191.1 AHW205.1 AHW206.1 AHW211.1 AHW213.1 AHW214.1 AHW230.1 AHW232.1 AHW240.1 AHW243.1 AHW251.1 AHW257.1 AHW261.1 AHW265.1 AHW277.1 AHW284.1 AHW310.1 AHW317.1 AHW329.1 AHW337.1 AHW341.1 AHW344.1 AHW345.1 AHW353.1 AHW358.1 AHW373.1 AHW377.1 AHW386.1 AHW389.1 AHW399.1 AHW400.1 AHW411.1 AHW419.1 AHW422.1 AHW423.1 AHW427.1 AHW439.1 AHW444.1 AHW448.1 AHW455.1 AHW458.1 AHW459.1 AHW460.1 AHW468.1 AHW469.1 AHW473.1 AHW478.1 AHW486.1 AHW509.1 AHW511.1 AHW516.1 AHW518.1 AHW520.1 AHW521.1 AHW524.1 AHW530.1 AHW536.1 AHW538.1 AHW539.1 AHW540.1 AHW549.1 AHW560.1 AHW566.1 AHW576.1 AHW597.1 AHW601.1 AHW617.1 AHW620.1 AHW622.1 AHW624.1 AHW634.1 AHW639.1 AHW644.1 AHW645.1 AHW649.1 AHW652.1 AJR080.1 AJW366.1 AJW459.1 AJW507.1 AJW511.1 AJW554.1 AJW558.1 AJW572.1 AJW580.1 AJW596.1 AJW607.1 AJW621.1 AJW623.1 AKB521.1 AKB527.1 AKB531.1 AKB537.1 AKB560.1 AKB565.1 AKB566.1 AKB573.1 AKB584.1 AKB595.1 AKB598.1 AKB612.1 AKB613.1 AKB614.1 AKB621.1 AKB633.1 AKB638.1 AKB640.1 AKB642.1 AKB650.1 AKB654.1 AKB675.1 AKB682.1 AKB683.1 AKB712.1 AKB714.1 AKB719.1 AKB724.1 AKB734.1 AKB736.1 AKB755.1 AKB756.1 AKB759.1 AKB762.1 AKB767.1 AKB776.1 AKB779.1 AKB780.1 AKB783.1 AKB785.1 AKB802.1 AKB806.1 AKB809.1 AKB810.1 AKB814.1 AKB825.1 AKB828.1 AKB832.1 AKB843.1 AKB848.1 AKB856.1 AKB864.1 AKB866.1 AKB869.1 AKB876.1 AKB881.1 AKB896.1 AKB897.1 AKB901.1 AKB903.1 AKB922.1 AKB940.1 AKB942.1 AKB951.1 AKB952.1 AKB956.1 AKB959.1 AKB962.1 AKB974.1 AKB979.1 AKB981.1 AKB985.1 AKB994.1 AKC002.1 AKC007.1 AKC010.1 AKC012.1 AKC022.1 AKC024.1 AKC026.1 AKC028.1 AKC049.1 AKC051.1 AKC055.1 AKC060.1 AKC076.1 AKC081.1 AKC086.1 AKC111.1 AKC112.1 AKC114.1 AKC119.1 AKC123.1 AKC126.1 AKC136.1 AKC144.1 AKC155.1 AKC159.1 AKC163.1 AKU509.1 AKU514.1 AKU522.1 AKU524.1 AKU525.1 AKU527.1 AKU530.1 AKU538.1 AKU550.1 AKU573.1 AKU574.1 AKU577.1 AKU578.1 AKU585.1 AKU593.1 AKU597.1 AKU606.1 AKU622.1 AKU625.1 AKU636.1 AKU638.1 AKU655.1 AKU656.1 AKU662.1 AKU668.1 AKU670.1 AKU680.1 AKU686.1 AKU692.1 AKU693.1 AKU695.1 AKU713.1 AKU714.1 AKU715.1 AKU716.1 AKU718.1 AKU722.1 AKU725.1 AKU731.1 AKU734.1 AKU737.1 AKU739.1 AKU753.1 AKU760.1 AKU763.1 AKU772.1 AKU777.1 AKU791.1 AKU796.1 AKU798.1 AKU801.1 AKU806.1 AKU809.1 AKU810.1 AKU814.1 AKU817.1 AKU836.1 AKU839.1 AKU844.1 AKU848.1 AKU851.1 AKU862.1 AKU887.1 AKU890.1 AKU891.1 AKU897.1 AKU911.1 AKU914.1 AKU915.1 AKU920.1 AKU925.1 AKU931.1 AKU937.1 AKU947.1 AKU964.1 AKU967.1 AKU971.1 AKU977.1 AKU986.1 AKU987.1 AKU991.1 AKU992.1 AKU993.1 AKV001.1 AKV003.1 AKV004.1 AKV009.1 AKV012.1 AKV016.1 AKV022.1 AKV036.1 AKV047.1 AKV050.1 AKV058.1 AKV059.1 AKV067.1 AKV077.1 AKV089.1 AKV090.1 AKV092.1 AKV094.1 AKV095.1 AKV096.1 AKV100.1 AKV101.1 AKV102.1 AKV106.1 AKV114.1 AKV120.1 AKV130.1 AKV135.1 AKV137.1 AKV144.1 AKV147.1 AKV148.1 AKV157.1 AKV163.1 AKV167.1 AKV168.1 AKV169.1 AKV170.1 AKV171.1 AKV173.1 AKV177.1 AKV188.1 AKV189.1 AKV199.1 AKV212.1 AKV220.1 AKV221.1 AKV225.1 AKV226.1 AKV228.1 AKV233.1 AKV240.1 AKV241.1 AKV242.1 AKV252.1 AKV255.1 AKV256.1 AKV261.1 AKV265.1 AKV267.1 AKV269.1 AKV270.1 AKV271.1 AKV274.1 AKV276.1 AKV277.1 AKV281.1 AKV284.1 AKV286.1 AKV287.1 AKV288.1 AKV289.1 AKV299.1 AKV300.1 AKV307.1 AKV311.1 AKV314.1 AKV318.1 AKV320.1 AKV322.1 AKV324.1 AKV330.1 AKV335.1 AKV337.1 AKV338.1 AKV350.1 AKV355.1 AKV357.1 AKV358.1 AKV361.1 AKV363.1 AKV370.1 AKV371.1 AKV372.1 AKV377.1 AKV386.1 AKV387.1 AKV391.1 AKV394.1 AKV396.1 AKV397.1 AKV399.1 AKV401.1 AKV403.1 AKV406.1 AKV407.1 AKV409.1 AKV411.1 AKV417.1 AKV421.1 AKV431.1 AKV440.1 AKV445.1 AKV447.1 AKV451.1 AKV458.1 AKV464.1 AKV466.1 AKV467.1 AKV468.1 AKV475.1 AKV486.1 AKV488.1 AKV489.1 AKV492.1 AKV507.1 AKV611.1 ALE754.1 ALE758.1 ALE766.1 ALE773.1 ALE941.1 ALE978.1 ALE983.1 ALE988.1 ALE994.1 ALF010.1 ALF014.1 ALF049.1 ALF063.1 ALF064.1 ALF077.1 ALF101.1 ALF111.1 ALF142.1 ALF153.1 ALF167.1 ALJ325.1 ALJ331.1 ALJ353.1 ALJ371.1 ALJ373.1 ALJ413.1 ALJ414.1 ALJ416.1 ALJ432.1 ALJ458.1 ALJ795.1 ALJ824.1 ALK002.1 ALK014.1 ALK032.1 ALK038.1 ALK051.1 ALK061.1 ALY954.1 ALY991.1 ALZ109.1 ALZ110.1 ALZ115.1 ALZ117.1 ALZ118.1 ALZ128.1 ALZ131.1 ALZ138.1 ALZ139.1 ALZ222.1 ALZ226.1 ALZ232.1 ALZ233.1 ALZ247.1 ALZ254.1 ALZ265.1 ALZ269.1 ALZ270.1 ALZ276.1 ALZ281.1 ALZ282.1 ALZ287.1 ALZ289.1 ALZ291.1 ALZ302.1 ALZ305.1 ALZ307.1 ALZ360.1 ALZ374.1 ALZ378.1 ALZ379.1 ALZ384.1 ALZ392.1 ALZ393.1 ALZ402.1 ALZ407.1 ALZ412.1 ALZ413.1 ALZ414.1 ALZ427.1 ALZ429.1 ALZ430.1 AMJ773.1 AMJ848.1 AMK014.1 AMN719.1 AMN757.1 AMN973.1 AMN975.1 AMN986.1 AMN988.1 AMN991.1 AMP024.1 AMP228.1 AMP270.1 AMP273.1 AMR918.1 AMR929.1 AMR937.1 AMR939.1 AMR943.1 AMR954.1 AMR956.1 AMR959.1 AMR961.1 AMR963.1 AMR975.1 AMR976.1 AMR979.1 AMR982.1 AMR983.1 AMR987.1 AMR989.1 AMS203.1 AMS257.1 AMS267.1 AMS269.1 AMS289.1 AMS292.1 AMS293.1 AMS296.1 AMS305.1 AMS312.1 AMS324.1 AMS329.1 AMS334.1 AMS339.1 AMS340.1 AMS344.1 AMS361.1 AMS363.1 AMS370.1 AMS371.1 AMS391.1 AMS440.1 AMS709.1 AMS778.1 AMS795.1 AMT002.1 AMT011.1 AMT052.1 AMT057.1 AMT068.1 AMT086.1 AMT087.1 AMT099.1 AMT101.1 AMT105.1 AMT111.1 AMT121.1 AMT124.1 AMT195.1 AMT255.1 AMT310.1 AMT338.1 AMT345.1 AMT348.1 AMT349.1 AMT355.1 AMT365.1 ANG890.1 ANG929.1 ANG967.1 ANG972.1 ANH010.1 ANH023.1 ANH042.1 ANH067.1 ANH095.1 ANH098.1 ANH108.1 ANH111.1 ANH114.1 ANH115.1 ANH121.1 ANH139.1 ANH152.1 ANH155.1 ANH157.1 ANH174.1 ANH188.1 ANH196.1 ANH197.1 ANH226.1 ANH231.1 ANH235.1 ANH239.1 ANH242.1 ANH254.1 ANH262.1 ANH264.1 ANH268.1 ANH270.1 ANH273.1 ANH281.1 ANH286.1 ANH394.1 ANH395.1 ANH396.1 ANH422.1 ANH423.1 ANH425.1 ANH430.1 ANH431.1 ANH437.1 ANH441.1 ANH445.1 ANH447.1 ANH458.1 ANH459.1 ANH463.1 ANH482.1 ANH483.1 ANH484.1 ANH502.1 ANH504.1 ANH506.1 ANH522.1 ANH534.1 ANH535.1 ANH558.1 ANH565.1 ANH579.1 ANH583.1 ANH584.1 ANH587.1 ANH588.1 ANH590.1 ANH591.1 ANH592.1 ANH609.1 ANH616.1 ANH621.1 ANH629.1 ANH640.1 ANH650.1 ANL380.1 ANL382.1 ANL385.1 ANL388.1 ANL389.1 ANL392.1 ANL393.1 ANL395.1 ANL396.1 ANL397.1 ANL398.1 ANL399.1 ANL402.1 ANL403.1 ANL404.1 ANL407.1 ANL408.1 ANL409.1 ANL410.1 ANL412.1 ANL413.1 ANL415.1 ANL418.1 ANL420.1 ANL422.1 ANL423.1 ANL424.1 ANL425.1 ANL426.1 ANL427.1 ANL428.1 ANL430.1 ANL431.1 ANL434.1 ANL435.1 ANL436.1 ANL437.1 ANL438.1 ANL452.1 ANL455.1 ANL460.1 ANL461.1 ANL462.1 ANL464.1 ANL468.1 ANL469.1 ANL474.1 ANL476.1 ANL480.1 ANL482.1 ANL491.1 ANL493.1 ANL497.1 ANL498.1 ANL500.1 ANL501.1 ANL502.1 ANL506.1 ANL510.1 ANL517.1 ANL518.1 ANL519.1 ANL521.1 ANL541.1 ANL558.1 ANL564.1 ANL590.1 ANL684.1 ANL686.1 ANL687.1 ANL691.1 ANL694.1 ANL703.1 ANL706.1 ANL721.1 ANL726.1 ANL732.1 ANL734.1 ANL739.1 ANL744.1 ANL781.1 ANL836.1 ANL842.1 ANL868.1 ANL910.1 ANM069.1 ANT277.1 ANT279.1 ANT283.1 ANT286.1 ANT287.1 ANT296.1 ANT297.1 ANT298.1 ANT300.1 ANT307.1 ANT320.1 ANT342.1 ANT383.1 ANT385.1 ANT404.1 ANT411.1 ANT439.1 ANT441.1 ANT451.1 ANT452.1 ANT461.1 ANT472.1 ANT476.1 ANT477.1 ANT478.1 ANT481.1 ANT484.1 ANT534.1 ANT557.1 ANT574.1 ANT618.1 ANT637.1 ANT666.1 ANT754.1 ANT867.1 ANT888.1 ANT896.1 ANT900.1 ANT908.1 ANT911.1 ANT913.1 ANU020.1 ANU033.1 ANU075.1 ANU077.1 ANU088.1 ANU141.1 ANU172.1 ANU179.1 ANU214.1 ANU217.1 ANU244.1 ANU271.1 ANU298.1 ANU300.1 ANU321.1 ANU329.1 ANU333.1 ANU334.1 ANU387.1 ANU391.1 ANU398.1 ANU401.1 APA828.1 APA829.1 APA830.1 APA833.1 APA843.1 APA852.1 APA863.1 APA872.1 APA874.1 APA876.1 APA879.1 APA888.1 APA890.1 APA905.1 APA906.1 APA910.1 APA915.1 APA922.1 APA927.1 APA937.1 APA944.1 APA949.1 APA966.1 APB004.1 APB031.1 APB038.1 APB040.1 APB045.1 APB046.1 APB056.1 APB068.1 APB070.1 APB071.1 APB080.1 APB081.1 APB088.1 APB091.1 APB096.1 APB104.1 APB106.1 APB109.1 APB111.1 APB112.1 APB116.1 APB117.1 APB153.1 APB164.1 APB199.1 APB203.1 APB206.1 APB208.1 APB210.1 APB219.1 APB224.1 APB232.1 APB240.1 APB242.1 APB260.1 APB276.1 APB297.1 APB304.1 APB321.1 APB323.1 APB324.1 APB334.1 APB335.1 APB344.1 APB350.1 APB355.1 APB359.1 APB364.1 APB375.1 APB383.1 APB388.1 APB390.1 APB392.1 APB398.1 APB403.1 APB412.1 APB415.1 APB441.1 APB446.1 APB476.1 APB478.1 APB482.1 APB499.1 APB525.1 APB528.1 APB532.1 APB541.1 APB546.1 APB547.1 APB555.1 APB556.1 APB565.1 APB595.1 APB603.1 APB607.1 APB608.1 APB618.1 APB619.1 APB620.1 APB634.1 APB640.1 APB642.1 APB658.1 APB677.1 APB678.1 APB685.1 APB694.1 APB695.1 APB708.1 APB709.1 APB713.1 APB723.1 APB726.1 APB751.1 APB753.1 APB764.1 APB765.1 APB766.1 APB768.1 APB780.1 APB801.1 APB802.1 APB805.1 APB827.1 APB836.1 APB840.1 APB844.1 APB852.1 APB862.1 APB870.1 APB873.1 APB877.1 APB893.1 APB894.1 APB902.1 APB911.1 APB916.1 APB928.1 APB929.1 APB930.1 APB931.1 APB934.1 APB938.1 APB939.1 APB941.1 APB970.1 APB983.1 APC015.1 APC017.1 APC028.1 APC032.1 APC037.1 APC041.1 APC058.1 APC060.1 APC061.1 APC068.1 APC071.1 APC075.1 APC110.1 APE452.1 APE492.1 APE504.1 APE516.1 APE517.1 APE522.1 APE538.1 APE547.1 APE550.1 APE551.1 APE567.1 APE569.1 APE575.1 APE577.1 APE583.1 APE584.1 APE590.1 APE591.1 APE630.1 APE640.1 APE642.1 APE650.1 APE663.1 APE693.1 APE708.1 APE709.1 APE714.1 APE727.1 APE734.1 APE740.1 APE742.1 APE745.1 APE762.1 APE792.1 APE805.1 APE813.1 APE894.1 APE899.1 APE923.1 APE953.1 APE960.1 APE961.1 APE983.1 APE990.1 APE991.1 APF002.1 APF005.1 APF006.1 APF007.1 APF011.1 APF029.1 APF033.1 APF114.1 APF160.1 APF192.1 APK895.1 APK900.1 APK901.1 APK924.1 APK928.1 APK955.1 APK956.1 APK963.1 APK974.1 APK978.1 APK979.1 APK990.1 APK993.1 APK996.1 APL039.1 APL045.1 APL059.1 APL072.1 APL086.1 APL100.1 APL103.1 APL114.1 APL146.1 APL154.1 APL159.1 APL177.1 APL194.1 APL210.1 APL211.1 APL216.1 APL217.1 APL248.1 APL250.1 APL254.1 APL261.1 APL262.1 APL269.1 APL275.1 APL276.1 APL279.1 APL291.1 APL293.1 APL303.1 APL305.1 APL306.1 APL307.1 APL327.1 APL330.1 APL334.1 APL344.1 APL348.1 APL349.1 APL350.1 APL361.1 APL362.1 APL363.1 APL364.1 APL367.1 APL373.1 APL383.1 APL384.1 APL389.1 APL401.1 APL420.1 APL423.1 APL426.1 APL428.1 APL431.1 APL444.1 APL450.1 APL451.1 APL452.1 APL453.1 APL462.1 APL486.1 APL491.1 APL492.1 APL496.1 APL529.1 APL547.1 APL549.1 APL565.1 APL576.1 APL590.1 APL592.1 APL601.1 APL602.1 APL613.1 APL621.1 APL622.1 APL630.1 APL637.1 APS036.1 APS039.1 APS053.1 APS067.1 APS068.1 APS070.1 APS071.1 APS084.1 APS095.1 APS104.1 APS119.1 APS125.1 APS126.1 APS140.1 APS148.1 APS151.1 APS153.1 APS154.1 APS155.1 APS202.1 APS236.1 APS237.1 APS238.1 APS262.1 APS266.1 APS278.1 APS281.1 APS285.1 APS314.1 APS325.1 APS326.1 APS327.1 APS345.1 APS347.1 APS351.1 APS357.1 APS361.1 APS363.1 APS365.1 APS373.1 APS375.1 APS400.1 APS424.1 APS436.1 APS447.1 APS448.1 APS457.1 APS469.1 APS501.1 APS514.1 APZ186.1 APZ192.1 APZ238.1 APZ251.1 APZ256.1 APZ276.1 APZ277.1 APZ300.1 APZ301.1 APZ323.1 APZ330.1 APZ350.1 APZ412.1 APZ456.1 APZ459.1 APZ560.1 APZ561.1 APZ564.1 APZ587.1 APZ594.1 APZ598.1 APZ608.1 APZ613.1 APZ614.1 APZ618.1 APZ638.1 APZ667.1 APZ670.1 APZ682.1 APZ697.1 APZ710.1 APZ721.1 APZ722.1 APZ758.1 APZ761.1 APZ776.1 APZ819.1 APZ826.1 APZ828.1 ARF828.1 ARF833.1 ARF858.1 ARF859.1 ARF860.1 ARF862.1 ARF867.1 ARF877.1 ARF880.1 ARF885.1 ARF889.1 ARF897.1 ARF949.1 ARF950.1 ARF953.1 ARF961.1 ARF964.1 ARF966.1 ARF970.1 ARF971.1 ARF976.1 ARF982.1 ARF993.1 ARG024.1 ARG028.1 ARG033.1 ARG037.1 ARG039.1 ARG044.1 ARG045.1 ARG047.1 ARG050.1 ARG056.1 ARG059.1 ARG068.1 ARG077.1 ARG123.1 ARG133.1 ARG140.1 ARG146.1 ARG156.1 ARG166.1 ARG168.1 ARG180.1 ARG186.1 ARG196.1 ARG201.1 ARG211.1 ARG214.1 ARG216.1 ARG222.1 ARG226.1 ARG227.1 ARG239.1 ARG254.1 ARG260.1 ARG264.1 ARG265.1 ARG266.1 ARG271.1 ARG273.1 ARG274.1 ARG276.1 ARG280.1 ARG282.1 ARG285.1 ARG297.1 ARG317.1 ARG321.1 ARG340.1 ARG344.1 ARG350.1 ARG368.1 ARG369.1 ARG373.1 ARG379.1 ARG382.1 ARG438.1 ARG442.1 ARG446.1 ARG447.1 ARG462.1 ARG477.1 ARG488.1 ARG490.1 ARG496.1 ARG501.1 ARG502.1 ARG503.1 ARG510.1 ARG512.1 ARG521.1 ARG522.1 ARG525.1 ARG526.1 ARG529.1 ARG535.1 ARG541.1 ARG542.1 ARG550.1 ARG634.1 ARG635.1 ARG639.1 ARG643.1 ARG645.1 ARG649.1 ARG650.1 ARG661.1 ARG662.1 ARJ277.1 ARK058.1 ARK064.1 ARK065.1 ARK092.1 ARK111.1 ARK144.1 ARK149.1 ARK166.1 ARK172.1 ARK188.1 ARK206.1 ARK218.1 ARK219.1 ARK222.1 ARK223.1 ARK224.1 ARK228.1 ARK242.1 ARK244.1 ARK247.1 ARK258.1 ARK261.1 ARK267.1 ARK277.1 ARK288.1 ARK301.1 ARK325.1 ARK335.1 ARK337.1 ARK347.1 ARK356.1 ARK367.1 ARK403.1 ARK404.1 ARK409.1 ARK411.1 ARK414.1 ARK416.1 ARK417.1 ARK435.1 ASJ075.1 ASM524.1 ASM531.1 ASM665.1 ASM687.1 ASM692.1 ASM765.1 ASR938.1 ASV349.1 ASV352.1 ASV357.1 ASV360.1 ASV370.1 ASV374.1 ASV412.1 ASV433.1 ASV462.1 ASV463.1 ASV603.1 ASV691.1 ASV693.1 ASV739.1 ASV746.1 ASV761.1 ASV844.1 ASV854.1 ASV875.1 ASV895.1 ASV942.1 ASV957.1 ASV959.1 ASW057.1 ASW084.1 YW0086.1 ZS0578.1 ZX5726.1 ZX5848.1
Worldwide Distribution - US (nationwide) (foreign consignee list will be provide by the firm at a later time)
1744 devices