FDA Enforcement Class II Ongoing

FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP System Analyzer

Recall: Z-0022-2025 · Reported October 16, 2024

Enforcement

Recall Number
Z-0022-2025
Event ID
95326
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Qualigen Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 16, 2024
Initiation Date
August 26, 2024
Classification Date
October 8, 2024
Address
2042 Corte Del Nogal, Ste B, Carlsbad, CA, 92011-1438, United States

Description

FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP System Analyzer

Reason

Thyroid Stimulating Hormone (TSH) calibrator card has an incorrect barcode linked to incorrect information, that may allow successful system calibration, but the TSH results obtained could be higher than expected.

Code Info

UDI-DI: 00816467020099, Lot: 2405017-1, Expiration: 2025-05-15

Distribution

US: TX, MS, NY, OH, MT, WA, PA, MN, SC, GA, NC, NV, WY, AK, SD, KS, NE, VT, AL, MA, OK, UT, WI, IA, WV. OUS: Germany, Switzerland

Quantity

60