FDA Enforcement Class II Terminated

ImplantDirect Legacy 3 Implant, REF 855216, SBM implant assembly, 2mmD x 16mmL, 4.5mmD platform, Rx, Sterile. The mislabeled model/part number on the vial cap is 835216. The responsible firm on the label is Implant Direct Sybron Manufacturing LLC, Thousand Oaks, CA. - Product Usage:Implant Direct Sybron Manufacturing LLC s dental implant product line consists of one-piece and two-piece implants for one-stage or two-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework.

Recall: Z-0018-2020 · Reported October 9, 2019

Enforcement

Recall Number
Z-0018-2020
Event ID
83518
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Implant Direct Sybron Manufacturing, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 9, 2019
Initiation Date
August 8, 2018
Classification Date
October 1, 2019
Termination Date
October 7, 2019
Address
3050 E Hillcrest Dr, N/A, Westlake Village, CA, 91362-3171, United States

Description

ImplantDirect Legacy 3 Implant, REF 855216, SBM implant assembly, 2mmD x 16mmL, 4.5mmD platform, Rx, Sterile. The mislabeled model/part number on the vial cap is 835216. The responsible firm on the label is Implant Direct Sybron Manufacturing LLC, Thousand Oaks, CA. - Product Usage:Implant Direct Sybron Manufacturing LLC s dental implant product line consists of one-piece and two-piece implants for one-stage or two-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework.

Reason

Two units in distribution were found to be mislabeled on the vial cap with the incorrect part number.

Code Info

Lot number 110350

Distribution

Worldwide distribution - US Nationwide including the states of CO, HI, ID, KS, NC, OH, OR, PA, SC, TN, and UT, and countries of Canada and the United Arab Emirates.

Quantity

22 vials