Getinge 88-Series Washer- Disinfector. Model Number: 88-5.
Enforcement
- Recall Number
- Z-0016-2026
- Event ID
- 97535
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Getinge Disinfection Ab
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 8, 2025
- Initiation Date
- August 29, 2025
- Classification Date
- October 1, 2025
- Address
- Ljungadalsgatan 11, N/A, Vaxjo, N/A, N/A, Sweden
Description
Getinge 88-Series Washer- Disinfector. Model Number: 88-5.
Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
Model Number: 88-5. UDI-DI Numbers: 07340153700109, 07340153700116. Software Version: GD 14080 Rev A. All Lot Numbers with electrical heating.
Worldwide - US Nationwide distribution in the states of Colorado, Massachusetts, New Mexico, New Hampshire and the countries of Algeria, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Austria, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Kenya, Korea, Republic of, Kuwait, Lithuania, Mexico, Netherlands, Norway, New Zealand, Poland, Qatar, Romania, Saudi Arabia, Singapore, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.
902 units