FDA Enforcement Class II Ongoing

CADD Medication Cassette Reservoir, 50mL, Product Code 21-7001-24

Recall: Z-0014-2025 · Reported October 30, 2024

Enforcement

Recall Number
Z-0014-2025
Event ID
95305
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Smiths Medical ASD Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 30, 2024
Initiation Date
June 26, 2024
Classification Date
October 18, 2024
Address
6000 Nathan Ln N, N/A, Minneapolis, MN, 55442-1690, United States

Description

CADD Medication Cassette Reservoir, 50mL, Product Code 21-7001-24

Reason

Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.

Code Info

UDI/DI 10610586022111, Lot Numbers: 4040237, 4013368, 3983342, 4046817, 405240, 4060903

Distribution

Worldwide distribution.

Quantity