FDA Enforcement
Class II
Terminated
Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009
Recall: Z-0014-2018
·
Reported October 18, 2017
Enforcement
- Recall Number
- Z-0014-2018
- Event ID
- 78130
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Handicare AB
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- October 18, 2017
- Initiation Date
- July 28, 2016
- Classification Date
- October 11, 2017
- Termination Date
- December 17, 2018
- Address
- Maskinvagen 17, Lulea, N/A, Sweden
Description
Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009
Reason
A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.
Code Info
a) Carina350EM, REF 6060011, Serial numbers: 661102000 - 661102160 b) Carina350EML, REF 60600012, Serial numbers: 661200500 - 661200584 c) Carina350EE, REF 60600009, Serial numbers: 660900500 -660900571
Distribution
OH, PA, MD
Quantity
9 units