FDA Enforcement Class II Terminated

Scan Abutment SP, Cat. No. 30-70101 - Product Usage: used to take an impression with an intra oral scanner or optical scanner to provide a digital registration of an implant or abutment platform. The Paltop Advanced Dental Solutions Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient s chewing function.

Recall: Z-0013-2022 · Reported October 13, 2021

Enforcement

Recall Number
Z-0013-2022
Event ID
88576
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Paltop Advanced Dental Solutions Ltd
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 13, 2021
Initiation Date
July 26, 2021
Classification Date
October 4, 2021
Termination Date
September 21, 2023
Address
5, Ha-Shita, Caesarea, N/A, Israel

Description

Scan Abutment SP, Cat. No. 30-70101 - Product Usage: used to take an impression with an intra oral scanner or optical scanner to provide a digital registration of an implant or abutment platform. The Paltop Advanced Dental Solutions Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient s chewing function.

Reason

Some devices are outside tolerances and showed a maximal deviation of 2 degrees in the orientation between the peek head and titanium base, and an incongruence between the size engraved on the peek has been identified. Deviations may delay the final restoration.

Code Info

All lots manufactured prior to July 21, 2021.

Distribution

US Nationwide distribution. No foreign distribution from US.

Quantity

782