FDA Enforcement Class II Ongoing

CADD Medication Cassette Reservoir, 100mL, FS, YELLOW, Product Code 21-7300-24

Recall: Z-0012-2025 · Reported October 30, 2024

Enforcement

Recall Number
Z-0012-2025
Event ID
95305
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Smiths Medical ASD Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 30, 2024
Initiation Date
June 26, 2024
Classification Date
October 18, 2024
Address
6000 Nathan Ln N, N/A, Minneapolis, MN, 55442-1690, United States

Description

CADD Medication Cassette Reservoir, 100mL, FS, YELLOW, Product Code 21-7300-24

Reason

Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.

Code Info

UDI/DI 10610586027192, Lot Numbers: 3971732, 3971731, 4034016, 3977457, 4013365, 4019421, 4019422, 4034015, 4037744, 4037745, 4034018, 4034017, 4046844, 4046843, 4053912, 4048899

Distribution

Worldwide distribution.

Quantity

26160 units