FDA Enforcement Class II Ongoing

CADD Medication Cassette Reservoir, 100mL, Product Code 21-7002-24

Recall: Z-0010-2025 · Reported October 30, 2024

Enforcement

Recall Number
Z-0010-2025
Event ID
95305
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Smiths Medical ASD Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 30, 2024
Initiation Date
June 26, 2024
Classification Date
October 18, 2024
Address
6000 Nathan Ln N, N/A, Minneapolis, MN, 55442-1690, United States

Description

CADD Medication Cassette Reservoir, 100mL, Product Code 21-7002-24

Reason

Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.

Code Info

UDI/DI 1061058602228, Lot Numbers: 3977441, 4013370, 3983314, 4013371, 4013369, 4013372, 4025948, 4033993, 4037444, 4040238

Distribution

Worldwide distribution.

Quantity

125928 units