FDA Enforcement Class II Ongoing

G7 Acetabular System, Acetabular Shell, Item numbers Item Numbers: 010000663 (52 MM E), 010000667 (60 MM G). used in hip arthroplasty

Recall: Z-0007-2023 · Reported October 12, 2022

Enforcement

Recall Number
Z-0007-2023
Event ID
90858
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Zimmer, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 12, 2022
Initiation Date
August 25, 2022
Classification Date
October 5, 2022
Address
1800 W Center St, N/A, Warsaw, IN, 46580-2304, United States

Description

G7 Acetabular System, Acetabular Shell, Item numbers Item Numbers: 010000663 (52 MM E), 010000667 (60 MM G). used in hip arthroplasty

Reason

The outer sterile package cavity has a corner wall thickness that is below the specification. The thin corner wall could potentially crack during transit. Potential risks include non-clinically or clinically significant extension of surgery, or infection leading to surgical intervention.

Code Info

Item No. 010000663 UDI-DI: 00880304524224 Lots: 7286477 7289929 7296576 7296633 7301697 Exp. 21-Jul-32 Item No. 010000667 UDI-DI: 00880304524262 Lot: 7253325 Exp. 13-Jun-32

Distribution

Domestic distribution US nationwide.

Quantity

56 US