FDA Enforcement Class II Ongoing

Miami J Select Collar, Catalog: MJS-101; Miami J Select Collar Set, Catalog: MJSR-101

Recall: Z-0006-2025 · Reported October 9, 2024

Enforcement

Recall Number
Z-0006-2025
Event ID
95278
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Ossur H / F
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 9, 2024
Initiation Date
August 20, 2024
Classification Date
October 1, 2024
Address
Grjothals 5, N/A, Reykjavik, N/A, N/A, Iceland

Description

Miami J Select Collar, Catalog: MJS-101; Miami J Select Collar Set, Catalog: MJSR-101

Reason

Pressure sensitive adhesive hooks attached to the sides of the anterior cervical orthosis collar panel can occasionally exhibit poor adhesion, which may lead to them gradually peeling off with repeated doffing, possibly leading to reduced immobilization of the cervical spine.

Code Info

REF/UDI-DI/GTIN/Lot Range/Sold in the Period, Date Range: MJS-101/05690967817008/MX220516 to MX230104/2022-05-19 to 2023-06-05, MJSR-101/05690967817145/MX220516 to MX230104/2022-05-19 to 2023-06-05

Distribution

US: NH, MA, NY, NJ, PA, GA, MD, NC, IN, FL, AR, TN, KY, OH, MI, IL, TX, MO, CO, UT, AZ, WA, OR, CA, NM, VA, IA, MN, NE, SC, MT, LA, ID, SD, AL, CT, OK, AK, PR

Quantity

21,020