Carto 3 EP Navigation System. Electro physiology system which views of the electrical activity of the heart through real-time data on 3-D, color-coded cardiac maps.
Enforcement
- Recall Number
- Z-0002-2016
- Event ID
- 72167
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Biosense Webster, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 7, 2015
- Initiation Date
- September 10, 2015
- Classification Date
- October 1, 2015
- Termination Date
- May 13, 2016
- Address
- 15715 Arrow Hwy, N/A, Irwindale, CA, 91706-2006, United States
Description
Carto 3 EP Navigation System. Electro physiology system which views of the electrical activity of the heart through real-time data on 3-D, color-coded cardiac maps.
Image disappeared from the cardiac ultrasound system when the CARTO 3 EP Navigation System needed restarting while the patient was experiencing pericardial effusion. Affects the CARTOSOUND Module of the CARTO 3 EP Navigation System when used with the SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. New precautions added.
Catalog #: C3SOUND, 10439236, 10439072, 10439011, 10438577
Worldwide Distribution. US nationwide, Italy, Canada, Japan, China, Singapore, Colombia, England, Russia, Spain, France, Germany, Finland, Ireland, Australia, Sweden, Hungary, India, Brazil, Slovenia, Netherlands, Mexico, and South Africa.
1035 (U.S.)