CDI¿ H/S Cuvette The intended use for the CDI¿ H IS Cuvette is as follows: The CDI¿ H/S Cuvettes are intended for use with a CDI monitor during cardiopulmonary bypass procedures when continuous monitoring of blood hematocrit and oxygen saturation is desired.
Enforcement
- Recall Number
- Z-0001-2017
- Event ID
- 75118
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Terumo Cardiovascular Systems Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- October 12, 2016
- Initiation Date
- December 11, 2015
- Classification Date
- October 3, 2016
- Termination Date
- October 26, 2016
- Address
- 125 Blue Ball Rd, N/A, Elkton, MD, 21921-5315, United States
Description
CDI¿ H/S Cuvette The intended use for the CDI¿ H IS Cuvette is as follows: The CDI¿ H/S Cuvettes are intended for use with a CDI monitor during cardiopulmonary bypass procedures when continuous monitoring of blood hematocrit and oxygen saturation is desired.
Terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette does not make a proper connection to the CDI H/S Probe. In this situation, blood parameter values for HCT, Hgb, and SO2 do not display.
6912 , 6913 6914, 6932, 6933 6934, 6922, 6923 6924, CV-6912, CV-6913, CV-6914 Cardiovascular Procedure Kits- Various
Worldwide Distribution - USA (nationwide) Distribution.
140,000 units