FDA Enforcement Class III Terminated

CDI¿ H/S Cuvette The intended use for the CDI¿ H IS Cuvette is as follows: The CDI¿ H/S Cuvettes are intended for use with a CDI monitor during cardiopulmonary bypass procedures when continuous monitoring of blood hematocrit and oxygen saturation is desired.

Recall: Z-0001-2017 · Reported October 12, 2016

Enforcement

Recall Number
Z-0001-2017
Event ID
75118
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Terumo Cardiovascular Systems Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
October 12, 2016
Initiation Date
December 11, 2015
Classification Date
October 3, 2016
Termination Date
October 26, 2016
Address
125 Blue Ball Rd, N/A, Elkton, MD, 21921-5315, United States

Description

CDI¿ H/S Cuvette The intended use for the CDI¿ H IS Cuvette is as follows: The CDI¿ H/S Cuvettes are intended for use with a CDI monitor during cardiopulmonary bypass procedures when continuous monitoring of blood hematocrit and oxygen saturation is desired.

Reason

Terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette does not make a proper connection to the CDI H/S Probe. In this situation, blood parameter values for HCT, Hgb, and SO2 do not display.

Code Info

6912 , 6913 6914, 6932, 6933 6934, 6922, 6923 6924, CV-6912, CV-6913, CV-6914 Cardiovascular Procedure Kits- Various

Distribution

Worldwide Distribution - USA (nationwide) Distribution.

Quantity

140,000 units