FDA Adverse Event Malfunction Summary report: N

EQUINOX

MDR report key: 9998499 · Received April 24, 2020

Report

Report Number
3004972304-2018-00036
Event Type
Malfunction
Date Received
April 24, 2020
Date of Event
August 3, 2018
Report Date
August 8, 2019
Manufacturer
CAIRE INC.
Product Code
CAW
PMA / PMN Number
K120785
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

"PURSUANT TO TITLE 21 - FOOD AND DRUGS, CHAPTER I - FOOD AND DRUG ADMINISTRATION DEPARTMENT OF HEALTH AND HUMAN SERVICES, SUBCHAPTER H -0 MEDICAL DEVICE, PART 803 - MEDICAL DEVICE REPORTING, SUBPART A - GENERAL PROVISIONS, SECTION 803.16, NEITHER THIS REPORT NOR ANY INFORMATION SUBMITTED HEREIN CONSTITUTES AN ADMISSION BY CAIRE INC. THAT THE DEVICE STATED IN THIS REPORT, CAIRE INC., OR CAIRE INC.'S EMPLOYEES, CAUSED OR CONTRIBUTED TO THE REPORTABLE EVENT STATED HEREIN." UNIT IS NOT AVAILABLE FOR TESTING. BASED ON THE DESCRIPTION OF THE INCIDENT, THE UNIT WAS ALARMING AS INTENDED. ALSO, THE UNIT IS NOT INTENDED FOR LIFE SUSTAINING PURPOSES.

Description of Event or Problem · 0

THIS REPORT WAS ORIGINALLY SUBMITTED ON 8/8/2019, AND IS BEING RESUBMITTED ON 7/6/2020 AS THE ORIGINAL SUBMISSION FAILED TO GO THROUGH. VITALITY MEDICAL HAS BEEN IN CONTACT WITH THE CUSTOMER'S DAUGHTER. OUR CUSTOMER JUST RECENTLY GOT IT BACK FROM THIS LAST REPAIR AND SAYS THAT THERE IS A YELLOW STEADY LIGHT THAT WON'T COME OFF, THE BATTERY IS FLASHING, THE DIAGNOSTIC COMES UP WITH AN ALARM 0100 AND THEN A RED LIGHT AND CODE FAIL 94 COMES ON AND THE UNIT SHUTS DOWN. ON THIS LAST ONE HER MOTHER WAS USING THE UNIT WHEN IT SHUT OFF AND SHE HAD TO RUSH TO GET AN OXYGEN TANK AS SHE IS LIFE DEPENDENT ON THE MACHINE.

Additional Manufacturer Narrative · 1

UNIT IS BEING RETURNED FOR EVALUATION. IF ANY NEW INFORMATION IS DISCOVERED, A FOLLOW UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THIS REPORT WAS ORIGINALLY SUBMITTED ON 9/4/2018, AND IS BEING RESUBMITTED ON 4/24/2020 AS THE ORIGINAL SUBMISSION FAILED TO GO THROUGH. VITALITY MEDICAL HAS BEEN IN CONTACT WITH THE CUSTOMER'S DAUGHTER. OUR CUSTOMER JUST RECENTLY GOT IT BACK FROM THIS LAST REPAIR AND SAYS THAT THERE IS A YELLOW STEADY LIGHT THAT WON'T COME OFF, THE BATTERY IS FLASHING, THE DIAGNOSTIC COMES UP WITH AN ALARM 0100 AND THEN A RED LIGHT AND CODE FAIL 94 COMES ON AND THE UNIT SHUTS DOWN. ON THIS LAST ONE HER MOTHER WAS USING THE UNIT WHEN IT SHUT OFF AND SHE HAD TO RUSH TO GET AN OXYGEN TANK AS SHE IS LIFE DEPENDENT ON THE MACHINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
459943 EQUINOX CONCENTRATOR, OXYGEN, TRANSPORTABLE CAW CAIRE INC. 4837-SEQ

Patients

Seq Age Sex Outcome Treatment
1